García-Resúa Carlos, Giráldez Fernández María Jesús, Yebra-Pimentel Eva, García-Montero Silvia
Department of Applied Physics (Optometry Group), University of Santiago de Compostela, Escuela de Optica y Optometria, Campus Sur, Santiago de Compostela - Spain.
Eur J Ophthalmol. 2010 May-Jun;20(3):523-30. doi: 10.1177/112067211002000326.
To assess the accuracy and repeatability of intraocular pressure (IOP) measurements obtained with the Canon TX-10 noncontact tonometer (NCT), using the Goldmann applanation tonometer (GAT) as the gold standard.
Seventy-three young healthy subjects were enrolled in the study. Central corneal thickness (CCT) was obtained with Orbscan Topography System and then repeated IOP measurements were obtained with both tonometers. Canon TX-10 NCT was performed first, followed by GAT. Another study sample of 32 young healthy subjects was used to test the repeatability with both tonometers. Two sets of data were acquired separated by 1 week. Plotting the difference between the methods against the mean was done to compare the tonometers and the intersession variability. The hypothesis of zero bias was examined by a paired t test. The 95% limits of agreement (LoA) were also calculated.
No statistically significant differences were found when comparing the Canon TX-10 NCT with GAT, displaying close level of agreement with GAT as seen by the 95% LoA (-4.78 mmHg to +4.00 mmHg). CCT ranged from 419 microm to 585 microm and no relationship was found between CCT and IOP measurements. The coefficients of repeatability were 3.70 mmHg and 3.41 mmHg for GAT and TX-10 tonometers, respectively.
There was a good agreement between Canon TX-10 and GAT in healthy subjects. From the results obtained here, CCT appears to not affect IOP measurements when CCT is thinner than 570 microm.
以Goldmann压平眼压计(GAT)作为金标准,评估使用佳能TX-10非接触眼压计(NCT)测量眼压(IOP)的准确性和可重复性。
73名年轻健康受试者参与本研究。使用Orbscan地形系统测量中央角膜厚度(CCT),然后用两种眼压计重复测量眼压。先进行佳能TX-10 NCT测量,随后进行GAT测量。另外选取32名年轻健康受试者作为研究样本,测试两种眼压计的可重复性。两组数据采集间隔1周。绘制两种方法测量结果的差值与平均值的关系图,以比较两种眼压计及测量结果在不同测量时段的变异性。采用配对t检验检验偏差为零的假设。还计算了95%一致性界限(LoA)。
将佳能TX-10 NCT与GAT比较时,未发现统计学上的显著差异,95% LoA显示其与GAT的一致性程度较高(-4.78 mmHg至+4.00 mmHg)。CCT范围为419微米至585微米,未发现CCT与眼压测量值之间存在相关性。GAT和TX-10眼压计的可重复性系数分别为3.70 mmHg和3.41 mmHg。
在健康受试者中,佳能TX-10与GAT之间具有良好的一致性。根据本研究结果,当CCT小于570微米时,CCT似乎不影响眼压测量值。