Bang Seung Pil, Lee Chong Eun, Kim Yu Cheol
Department of Ophthalmology, Keimyung University School of Medicine, Dongsan Medical Center, #56, Dalseong-ro, Jung-gu, Daegu, 41931, South Korea.
BMC Ophthalmol. 2017 Nov 2;17(1):199. doi: 10.1186/s12886-017-0593-1.
To compare the measurement of intraocular pressure (IOP) among the three different non-contact tonometers (NCT) and the Goldmann applanation tonometer (GAT) for non-glaucomatous subjects.
In 52 eyes of 52 non-glaucomatous subjects, IOP was measured sequentially with the Canon TX-20P, the Nidek NT-530P, the Topcon CT-1P, and the GAT at the same time. We evaluated the IOP-measurement agreement among the tonometers as well as the factors affecting the measurements.
A significant positive correlation was shown between the IOP values obtained with GAT and each NCT. The Canon TX-20P showed statistically the most significant agreement with the GAT (ICC 0.906, 95% CI 0.837-0.946). In an analysis of the Bland-Altman plots, the Canon TX-20P also showed the largest mean bias (1.38 mmHg) but the narrowest limits of agreement (LoA) (95% LoA; ± 3.43 mmHg). The Topcon CT-1P showed the smallest mean bias (0.48 mmHg) but the widest LoA (95% LoA; ± 4.16 mmHg). The Topcon CT-1P and Nidek NT-530P both showed a significantly positive correlation between increasing central corneal thickness (CCT) and increasing IOP.
There was a statistically significant correlation between each of the three different NCT and the GAT measurements. IOP measured with the Canon TX-20P and Topcon CT-1P tended to be higher, and with the Nidek NT-530P lower, than with the GAT. Practitioners need to know the properties of their own NCTs and their respective measurement tendencies.
比较三种不同非接触眼压计(NCT)与戈德曼压平眼压计(GAT)在非青光眼受试者中测量眼压(IOP)的情况。
对52例非青光眼受试者的52只眼,同时依次使用佳能TX - 20P、尼德克NT - 530P、拓普康CT - 1P和GAT测量眼压。我们评估了眼压计之间眼压测量的一致性以及影响测量的因素。
GAT测得的眼压值与各NCT测得的眼压值之间呈现显著正相关。佳能TX - 20P与GAT在统计学上显示出最显著的一致性(组内相关系数0.906,95%可信区间0.837 - 0.946)。在布兰德 - 奥特曼图分析中,佳能TX - 20P也显示出最大的平均偏差(1.38 mmHg),但一致性界限(LoA)最窄(95% LoA;± 3.43 mmHg)。拓普康CT - 1P显示出最小的平均偏差(0.48 mmHg),但LoA最宽(95% LoA;± 4.16 mmHg)。拓普康CT - 1P和尼德克NT - 530P均显示中央角膜厚度(CCT)增加与眼压升高之间存在显著正相关。
三种不同的NCT与GAT测量结果之间在统计学上存在显著相关性。与GAT相比,用佳能TX - 20P和拓普康CT - 1P测量的眼压往往较高,而用尼德克NT - 530P测量的眼压较低。从业者需要了解自己所使用NCT的特性及其各自的测量倾向。