Comparative Assessment of The Goldmann Applanation and Noncontact Tonometers in Intraocular Pressure Measurements in a Sample of Glaucoma Patients in the Cape Coast Metropolis, Ghana.

作者信息

Kyei Samuel, Gboglu Cynthia Pakyennu, Kwarteng Michael Agyemang, Assiamah Frank

机构信息

Department of Optometry and Vision Science, School of Allied Health Sciences, College of Health and Allied Sciences, University of Cape Coast, Cape Coast, Ghana.

Department of Optometry, Faculty of Science and Engineering, Bindura University of Science Education, Bindura, Zimbabwe.

出版信息

Niger Med J. 2020 Nov-Dec;61(6):323-327. doi: 10.4103/nmj.NMJ_177_20. Epub 2020 Dec 19.

Abstract

OBJECTIVES

The objective of the study was to explore the usage of the Goldmann applanation tonometry and noncontact tonometry interchangeably in the measurement of intraocular pressure (IOP) in glaucoma patients.

MATERIALS AND METHODS

The study involved 441 clinically diagnosed glaucoma patients receiving care at a referral facility. IOP measurements were obtained using both the Noncontact tonometer and Goldmann applanation tonometer The repeatability of the measures was analyzed by comparing the repeated measures of the devices using paired -test and calculating the correlation coefficient. A Bland-Altman analysis was used to determine the limits of agreement between the two procedures.

RESULTS

There were 271 (61.5%) males and 170 (38.5%) females and their age ranged from 18 to 73 years (mean age = 49.37; standard deviation ± 14.81 years). The findings of the study showed significantly lower readings ( < 0.001) of the GAT (right eye = 17.40 ± 7.48 mmHg; left eye = 16.80 ± 7.49 mmHg) compared to the NCT (right eye = 20.15 ± 8.30 mmHg; left eye = 19.74 ± 8.31 mmHg). There was a strong positive correlation between the GAT and NCT findings in the right eye ( = 0.871, = 441, < 0.001) and in the left eye ( = 0.887, = 441, < 0.001). There was a wide limit of agreement between NCT and GAT measurements.

CONCLUSION

There was statistically significant higher measures obtained with NCT than the GAT but did not exceed the allowable inter-device difference. There was a strong positive correlation between GAT and NCT measurements. However, it is strongly recommended that these devices are not used interchangeably in the monitoring of IOP in glaucoma due to the wide range of limits of agreement.

摘要

目的

本研究的目的是探讨在青光眼患者眼压(IOP)测量中,戈德曼压平眼压计和非接触眼压计能否交替使用。

材料与方法

本研究纳入了441名在转诊机构接受治疗的临床诊断青光眼患者。使用非接触眼压计和戈德曼压平眼压计测量眼压。通过配对检验比较两种设备的重复测量值并计算相关系数,分析测量的重复性。采用布兰德-奥特曼分析来确定两种测量方法之间的一致性界限。

结果

有271名(61.5%)男性和170名(38.5%)女性,年龄范围为18至73岁(平均年龄=49.37岁;标准差±14.81岁)。研究结果显示,与非接触眼压计(右眼=20.15±8.30 mmHg;左眼=19.74±8.31 mmHg)相比,戈德曼压平眼压计的读数显著更低(<0.001)(右眼=17.40±7.48 mmHg;左眼=16.80±7.49 mmHg)。戈德曼压平眼压计和非接触眼压计在右眼(=0.871,=441,<0.001)和左眼(=0.887,=441,<0.001)的测量结果之间存在强正相关。非接触眼压计和戈德曼压平眼压计测量之间的一致性界限较宽。

结论

非接触眼压计测量得到的数值在统计学上显著高于戈德曼压平眼压计,但未超过设备间允许的差异。戈德曼压平眼压计和非接触眼压计测量之间存在强正相关。然而,由于一致性界限范围较宽,强烈建议在青光眼眼压监测中不要交替使用这些设备。