Bijster P
Clinical Laboratory, Martini Ziekenhuis, Groningen, The Netherlands.
Eur J Clin Chem Clin Biochem. 1993 Mar;31(3):173-8.
A multi-centre study was undertaken to evaluate the performance of the Reflotron high density lipoprotein cholesterol method in twelve different clinical chemistry hospital laboratories. Results were compared with phosphotungistic acid/Mg2+ and polyethylene glycol/dextran sulphate/Mg2+ precipitation methods, routinely performed in these laboratories. Precision was found to be excellent, for both the day-to-day results (SD = 0.03 mmol/l) and for the overall between-laboratory precision (SD = 0.02 mmol/l), measured with control sera. HDL cholesterol concentrations measured with the Reflotron system were about 10% lower than those obtained with the two routine precipitation methods, using different instruments with different cholesterol standardization procedures.
一项多中心研究在12家不同的临床化学医院实验室中开展,以评估Reflotron高密度脂蛋白胆固醇检测方法的性能。将结果与这些实验室常规采用的磷钨酸/Mg2+和聚乙二醇/硫酸葡聚糖/Mg2+沉淀法进行比较。使用对照血清测量发现,无论是日常结果(标准差 = 0.03 mmol/l)还是实验室间总体精密度(标准差 = 0.02 mmol/l),精密度都非常出色。使用Reflotron系统测量的高密度脂蛋白胆固醇浓度,比采用两种常规沉淀法、使用不同仪器及不同胆固醇标准化程序所获得的浓度低约10%。