Abstract

AIMS

Using compartment modelling, we assessed the time delay between blood glucose and sensor glucose measured by the Guardian RT continuous glucose monitoring system in young subjects with Type 1 diabetes (T1D).

METHODS

Twelve children and adolescents with T1D treated by continuous subcutaneous insulin infusion (male/female 7/5; age 13.1 +/- 4.2 years; body mass index 21.9 +/- 4.3 kg/m(2); mean +/- sd) were studied over 19 h in a Clinical Research Facility. Guardian RT was calibrated every 6 h and sensor glucose measured every 5 min. Reference blood glucose was measured every 15 min using a YSI 2300 STAT Plus Analyser. A population compartment model of sensor glucose-blood glucose kinetics was adopted to estimate the time delay, the calibration scale and the calibration shift.

RESULTS

The population median of the time delay was 15.8 (interquartile range 15.2, 16.5) min, which was corroborated by correlation analysis between blood glucose and 15-min delayed sensor glucose. The delay has a relatively low intersubject variability, with 95% of individuals predicted to have delays between 10.4 and 24.3 min. Population medians (interquartile range) for the scale and shift are 0.800 (0.777, 0.823) (unitless) and 1.66 (1.47, 1.84) mmol/l, respectively.

CONCLUSIONS

In young subjects with T1D, the total time delay associated with the Guardian RT system was approximately 15 min. This is twice that expected on physiological grounds, suggesting a 5- to 10-min delay because of data processing. Delays above 25 min are rarely to be observed.

摘要

目的

使用房室模型,我们评估了 1 型糖尿病(T1D)年轻患者使用 Guardian RT 连续血糖监测系统测量的血糖与传感器血糖之间的时间延迟。

方法

12 名接受持续皮下胰岛素输注治疗的 T1D 儿童和青少年(男/女 7/5;年龄 13.1 +/- 4.2 岁;体重指数 21.9 +/- 4.3 kg/m(2);平均值 +/- sd)在临床研究设施中进行了 19 小时的研究。Guardian RT 每 6 小时校准一次,传感器血糖每 5 分钟测量一次。使用 YSI 2300 STAT Plus 分析仪每 15 分钟测量一次参考血糖。采用传感器血糖-血糖动力学的群体房室模型来估计时间延迟、校准比例和校准偏移。

结果

人群中位数的时间延迟为 15.8(四分位距 15.2,16.5)分钟,这与血糖与 15 分钟延迟的传感器血糖之间的相关分析相符。延迟具有相对较低的个体间变异性,95%的个体预计延迟在 10.4 至 24.3 分钟之间。人群中位数(四分位距)的比例和偏移分别为 0.800(0.777,0.823)(无单位)和 1.66(1.47,1.84)mmol/l。

结论

在 T1D 的年轻患者中,Guardian RT 系统相关的总时间延迟约为 15 分钟。这是基于生理原因预期的两倍,这表明由于数据处理会有 5-10 分钟的延迟。很少观察到超过 25 分钟的延迟。