Twin Research and Genetic Epidemiology Unit, King's College London School of Medicine and St. Thomas' Hospital Campus, London, UK.
Eye (Lond). 2010 Jul;24(7):1165-70. doi: 10.1038/eye.2010.11. Epub 2010 Feb 12.
The purpose of this study was to compare the reliability of the 'gold standard' Goldmann applanation tonometer (GAT), with that of the ocular response analyser (ORA), and the dynamic contour tonometer (DCT).
A total of 694 subjects were recruited to participate from the TwinsUK (UK Adult Twin Registry) at St Thomas' Hospital, London. Intraocular pressure (IOP) was measured using GAT, ORA, and the DCT. The agreement between the three methods was assessed using the Bland-Altman method. Repeatability coefficients and coefficient of variation between first and second readings of the same eye were used to assess reliability.
Mean age was 57.5 years (SD, 13.1; range, 16.1-88.5). The mean IOPs, calculated using the mean of two readings from the right eye were as follows: Goldmann (GAT), 14.1+/-2.8 mm Hg; IOPg (ORA), 15.9+/-3.2 mm Hg; IOPcc (ORA), 16.6+/-3.2 mm Hg; and DCT, 16.9+/-2.7 mm Hg. The 95% limits of agreement were for ORA (IOPcc): GAT, -2.07 to 7.18 mm Hg; for DCT: GAT, -0.49 to 6.21 mm Hg; and for DCT: ORA (IOPcc), -3.01 to 4.85 mm Hg. Coefficients of variation for the three tonometers were GAT, 8.3%; ORA, 8.2%; DCT, 6.3%. The repeatability coefficients were 3.4 mm Hg for GAT, 3.57 mm Hg for ORA and 3.09 mm Hg for DCT. GAT and ORA (IOPg) readings showed a positive correlation with central corneal thickness (P<0.005).
This study found similar reliability in all three tonometers. Bland-Altman plots showed the three instruments to have 95% limits of agreement outside the generally accepted limits, which means they are not interchangeable. GAT measurements were found to be significantly lower than the two newer instruments.
本研究旨在比较“金标准”压平眼压计(GAT)与眼反应分析仪(ORA)和动态轮廓眼压计(DCT)的可靠性。
共有 694 名受试者从伦敦托马斯医院的英国双胞胎协会(UK 成人双胞胎登记处)招募参与研究。使用 GAT、ORA 和 DCT 测量眼内压(IOP)。使用 Bland-Altman 法评估三种方法之间的一致性。使用同一眼的两次读数的重复性系数和变异系数来评估可靠性。
平均年龄为 57.5 岁(标准差,13.1;范围,16.1-88.5)。右眼两次读数平均值的平均 IOP 如下:GAT,14.1+/-2.8mmHg;IOPg(ORA),15.9+/-3.2mmHg;IOPcc(ORA),16.6+/-3.2mmHg;和 DCT,16.9+/-2.7mmHg。ORA(IOPcc)的 95%一致性界限为:GAT,-2.07 至 7.18mmHg;DCT:GAT,-0.49 至 6.21mmHg;和 DCT:ORA(IOPcc),-3.01 至 4.85mmHg。三种眼压计的变异系数分别为 GAT,8.3%;ORA,8.2%;DCT,6.3%。GAT 的重复性系数为 3.4mmHg,ORA 为 3.57mmHg,DCT 为 3.09mmHg。GAT 和 ORA(IOPg)读数与中央角膜厚度呈正相关(P<0.005)。
本研究发现三种眼压计具有相似的可靠性。Bland-Altman 图显示三种仪器的 95%一致性界限超出了普遍接受的范围,这意味着它们不能互换。GAT 测量值明显低于两种较新的仪器。