动态轮廓眼压测量与非接触眼压测量在高眼压症和青光眼中的比较。

Comparison of dynamic contour tonometry and noncontact tonometry in ocular hypertension and glaucoma.

机构信息

Department of Ophthalmology, Fatih University Medical School, Ankara, Turkey.

出版信息

Eye (Lond). 2009 Mar;23(3):663-8. doi: 10.1038/eye.2008.3. Epub 2008 Feb 8.

Abstract

PURPOSE

To assess the agreement in the measurement of intraocular pressure obtained by dynamic contour tonometer (DCT) and noncontact tonometer (NCT) in patients with glaucoma and ocular hypertension, to investigate the effect of corneal thickness on pressure readings by both instruments, and to assess the reproducibility of dynamic contour tonometer.

METHODS

NCT and DCT measurements were made on 104 eyes of 104 patients with primary open-angle glaucoma (n=75) or ocular hypertension (n=29), and agreement was assessed by means of Bland-Altman plots. The effect of corneal thickness on both tonometers was assessed by linear regression analysis. Interobserver and intraobserver variations for dynamic contour tonometer were assessed in 41 eyes of 41 patients.

RESULTS

The mean difference+/-SD (95% limits of agreement) between NCT and DCT was -0.80+/-2.98 (-6.6 to 5.1) mm Hg (P=0.009) and no relation between NCT/DCT differences and average was found. The intraocular pressure readings obtained by noncontact tonometer depended on central corneal thickness (P<0.001, adjusted r(2)=0.301). However, dynamic contour tonometer readings showed no effect of corneal thickness (P=0.388, adjusted r(2)=-0.002). The coefficient of repeatability for DCT was 0.92 (95% CI 0.85-0.96, P=0.001).

CONCLUSION

In subjects with primary open-angle glaucoma and ocular hypertension, NCT and DCT readings are not interchangeable. DCT measurements, unlike NCT measurements, did not depend on corneal thickness.

摘要

目的

评估在青光眼和高眼压症患者中使用动态轮廓眼压计(DCT)和非接触眼压计(NCT)测量眼压的一致性,研究角膜厚度对两种仪器眼压读数的影响,并评估动态轮廓眼压计的可重复性。

方法

对 104 例原发性开角型青光眼(n=75)或高眼压症(n=29)患者的 104 只眼进行 NCT 和 DCT 测量,并通过 Bland-Altman 图评估一致性。线性回归分析评估角膜厚度对两种眼压计的影响。在 41 例患者的 41 只眼中评估动态轮廓眼压计的观察者间和观察者内变异。

结果

NCT 和 DCT 之间的平均差异+/-标准差(95%一致性区间)为-0.80+/-2.98(-6.6 至 5.1)mmHg(P=0.009),且未发现 NCT/DCT 差异与平均值之间存在关系。非接触眼压计的眼压读数取决于中央角膜厚度(P<0.001,调整 r(2)=0.301)。然而,动态轮廓眼压计读数不受角膜厚度的影响(P=0.388,调整 r(2)=-0.002)。DCT 的可重复性系数为 0.92(95%置信区间 0.85-0.96,P=0.001)。

结论

在原发性开角型青光眼和高眼压症患者中,NCT 和 DCT 的读数不能互换。与 NCT 测量不同,DCT 测量不受角膜厚度的影响。