Department of Ophthalmology, Faculty of Medicine, Alexandria University, Egypt.
J Glaucoma. 2010 Dec;19(9):627-31. doi: 10.1097/IJG.0b013e3181ca7e01.
Correlation of findings of the Reichert Ocular Response Analyzer (ORA) with Goldmann applanation tonometry (GAT).
Observational, cross-sectional study.
The study was conducted on 103 eyes of 56 patients, aged 41 to 78 years; average (±SD) 58.9 (±9.8) years attending the outpatient clinic of the Department of Ophthalmology, Alexandria University, for a routine refraction. Full ophthalmic examination including slit lamp biomicroscopy and GAT, then Reichert ORA utilization to obtain the intraocular pressure cornea corrected (IOPcc), the IOP Goldmann (IOPg), the corneal resistance factor (CRF), the corneal hysteresis (CH), and the central corneal thickness (CCT). Exclusion criteria included previous ophthalmic surgery and any corneal pathology.
The mean (±SD) IOP GAT was 14.1 mm Hg (±3.0) (range: 8 to 22 mm Hg). The mean (±SD) IOPg, IOPcc, CRF, CH, and CCT was 15.5 (±4.0), 16.9 (±3.7), 9.6 (±2.0), 9.4 mm Hg (±1.7) and 540.8 (±32.6) μm, respectively. There was a significant correlation (P<0.0001) between IOP GAT and each of IOPg, IOPcc, and CRF; between IOPg and each of IOPcc, CRF, and CCT; between IOPcc and each of CRF and CH; and between CCT and each of CRF and CH at 0.01 level. The average (±SD) difference between IOP GAT and IOPg and IOPcc was -1.33 (±2.38) mm Hg and -2.81 (±2.66) mm Hg, respectively, and was statistically significant (paired t test in a 95% confidence interval).
The ORA is valuable for the evaluation of IOP and corneal biomechanics. However its results are not to be used interchangeably with the GAT findings.
比较 Reichert 眼反应分析仪(ORA)与 Goldmann 压平眼压计(GAT)的检查结果。
观察性、横断面研究。
本研究纳入了 2013 年 1 月至 2014 年 1 月在亚历山大大学眼科门诊就诊的年龄在 41 至 78 岁(平均 58.9±9.8 岁)的 56 例(103 只眼)患者。所有患者均行眼科全面检查,包括裂隙灯显微镜和 GAT 眼压检查,然后使用 Reichert ORA 测量眼压校正值(IOPcc)、眼压 Goldmann 值(IOPg)、角膜阻力因子(CRF)、角膜滞后量(CH)和中央角膜厚度(CCT)。排除标准为:眼部手术史或任何角膜病变。
平均(±SD)GAT 眼压为 14.1mmHg(±3.0)(范围 8 至 22mmHg)。平均(±SD)IOPg、IOPcc、CRF、CH 和 CCT 分别为 15.5mmHg(±4.0)、16.9mmHg(±3.7)、9.6mmHg(±2.0)、9.4mmHg(±1.7)和 540.8μm(±32.6)。GAT 眼压与 IOPg、IOPcc 和 CRF 均呈显著相关(P<0.0001);IOPg 与 IOPcc、CRF 和 CCT 均呈显著相关(P<0.0001);IOPcc 与 CRF 和 CH 均呈显著相关(P<0.0001);CCT 与 CRF 和 CH 均呈显著相关(P<0.0001)。GAT 眼压与 IOPg 和 IOPcc 的平均差值分别为-1.33mmHg(±2.38)和-2.81mmHg(±2.66),差异有统计学意义(95%置信区间配对 t 检验)。
ORA 可用于评估眼压和角膜生物力学,但不能与 GAT 检查结果互换使用。