Zareei Athar, Razeghinejad Mohammad Reza, Nowroozzadeh Mohammad Hosein, Mehrabi Yadollah, Aghazadeh-Amiri Mohammad
Department of Optometry, International Branch, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
Poostchi Ophthalmology Research Center, Department of Ophthalmology, Shiraz University of Medical Sciences, Shiraz, Iran.
J Ophthalmic Vis Res. 2015 Jan-Mar;10(1):43-8. doi: 10.4103/2008-322X.156105.
To determine the agreement between intraocular pressure (IOP) measurements using an automated non-contact tonometer (NCT), Goldmann applanation tonometer (GAT), and the ocular response analyzer (ORA) in subjects with primary congenital glaucoma (PCG).
Twenty-nine eyes of 17 PCG patients underwent IOP measurements using NCT, GAT and ORA. Variables obtained by the ORA were corneal-compensated IOP (IOPcc), Goldmann-correlated IOP (IOPg), corneal hysteresis (CH), and corneal resistance factor (CRF). A difference more than 1.5 mmHg for IOP was considered as clinically relevant.
Mean age of the patients was 12 years. Mean IOP (±standard deviation, SD) was 15.3 ± 2.8 mmHg (GAT), 15.5 ± 6.0 (NCT), 19.2 ± 7.0 (IOPg), and 21.1 ± 7.9 (IOPcc); (P = 0.001). Except for NCT vs. GAT (P = 1.0), the average IOP difference between each pair of measurements was clinically relevant. The 95% limits of agreements were - 10.2 to 10.3 mmHg (NCT vs. GAT), -7.8 to 15.3 (IOPg vs. GAT), and - 8.1 to 19.0 (IOPcc vs. GAT). The differences in IOP measurements increased significantly with higher average IOP values (r = 0.715, P = 0.001, for NCT vs. GAT; r = 0.802, P < 0.001, for IOPg vs. GAT; and r = 0.806, P < 0.001, for IOPcc vs. GAT). CH showed a significant association with differences in IOP measurements only for IOPcc vs. GAT (r = 0.830, P < 0.001).
Mean IOP obtained by NCT was not significantly different from that of GAT, but ORA measured IOPs were significantly higher than both other devices.
确定在原发性先天性青光眼(PCG)患者中使用自动非接触眼压计(NCT)、Goldmann压平眼压计(GAT)和眼反应分析仪(ORA)测量眼压(IOP)之间的一致性。
对17例PCG患者的29只眼使用NCT、GAT和ORA测量眼压。ORA获得的变量包括角膜补偿眼压(IOPcc)、Goldmann相关眼压(IOPg)、角膜滞后(CH)和角膜阻力因子(CRF)。眼压差异超过1.5 mmHg被认为具有临床相关性。
患者的平均年龄为12岁。平均眼压(±标准差,SD)分别为15.3±2.8 mmHg(GAT)、15.5±6.0(NCT)、19.2±7.0(IOPg)和21.1±7.9(IOPcc);(P = 0.001)。除了NCT与GAT(P = 1.0)外,每对测量之间的平均眼压差异均具有临床相关性。95%的一致性界限为-10.2至10.3 mmHg(NCT与GAT)、-7.8至15.3(IOPg与GAT)和-8.1至19.0(IOPcc与GAT)。随着平均眼压值升高,眼压测量的差异显著增加(NCT与GAT,r = 0.715,P = 0.001;IOPg与GAT,r = 0.802,P < 0.001;IOPcc与GAT,r = 0.806,P < 0.001)。仅在IOPcc与GAT之间,CH与眼压测量差异显示出显著相关性(r = 0.830,P < 0.001)。
NCT测得的平均眼压与GAT无显著差异,但ORA测得的眼压显著高于其他两种设备。