Intraocular pressure measurement by three different tonometers in primary congenital glaucoma.

作者信息

Zareei Athar, Razeghinejad Mohammad Reza, Nowroozzadeh Mohammad Hosein, Mehrabi Yadollah, Aghazadeh-Amiri Mohammad

机构信息

Department of Optometry, International Branch, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

Poostchi Ophthalmology Research Center, Department of Ophthalmology, Shiraz University of Medical Sciences, Shiraz, Iran.

出版信息

J Ophthalmic Vis Res. 2015 Jan-Mar;10(1):43-8. doi: 10.4103/2008-322X.156105.

Abstract

PURPOSE

To determine the agreement between intraocular pressure (IOP) measurements using an automated non-contact tonometer (NCT), Goldmann applanation tonometer (GAT), and the ocular response analyzer (ORA) in subjects with primary congenital glaucoma (PCG).

METHODS

Twenty-nine eyes of 17 PCG patients underwent IOP measurements using NCT, GAT and ORA. Variables obtained by the ORA were corneal-compensated IOP (IOPcc), Goldmann-correlated IOP (IOPg), corneal hysteresis (CH), and corneal resistance factor (CRF). A difference more than 1.5 mmHg for IOP was considered as clinically relevant.

RESULTS

Mean age of the patients was 12 years. Mean IOP (±standard deviation, SD) was 15.3 ± 2.8 mmHg (GAT), 15.5 ± 6.0 (NCT), 19.2 ± 7.0 (IOPg), and 21.1 ± 7.9 (IOPcc); (P = 0.001). Except for NCT vs. GAT (P = 1.0), the average IOP difference between each pair of measurements was clinically relevant. The 95% limits of agreements were - 10.2 to 10.3 mmHg (NCT vs. GAT), -7.8 to 15.3 (IOPg vs. GAT), and - 8.1 to 19.0 (IOPcc vs. GAT). The differences in IOP measurements increased significantly with higher average IOP values (r = 0.715, P = 0.001, for NCT vs. GAT; r = 0.802, P < 0.001, for IOPg vs. GAT; and r = 0.806, P < 0.001, for IOPcc vs. GAT). CH showed a significant association with differences in IOP measurements only for IOPcc vs. GAT (r = 0.830, P < 0.001).

CONCLUSION

Mean IOP obtained by NCT was not significantly different from that of GAT, but ORA measured IOPs were significantly higher than both other devices.

摘要

目的

确定在原发性先天性青光眼(PCG)患者中使用自动非接触眼压计(NCT)、Goldmann压平眼压计(GAT)和眼反应分析仪(ORA)测量眼压(IOP)之间的一致性。

方法

对17例PCG患者的29只眼使用NCT、GAT和ORA测量眼压。ORA获得的变量包括角膜补偿眼压(IOPcc)、Goldmann相关眼压(IOPg)、角膜滞后(CH)和角膜阻力因子(CRF)。眼压差异超过1.5 mmHg被认为具有临床相关性。

结果

患者的平均年龄为12岁。平均眼压(±标准差,SD)分别为15.3±2.8 mmHg(GAT)、15.5±6.0(NCT)、19.2±7.0(IOPg)和21.1±7.9(IOPcc);(P = 0.001)。除了NCT与GAT(P = 1.0)外,每对测量之间的平均眼压差异均具有临床相关性。95%的一致性界限为-10.2至10.3 mmHg(NCT与GAT)、-7.8至15.3(IOPg与GAT)和-8.1至19.0(IOPcc与GAT)。随着平均眼压值升高,眼压测量的差异显著增加(NCT与GAT,r = 0.715,P = 0.001;IOPg与GAT,r = 0.802,P < 0.001;IOPcc与GAT,r = 0.806,P < 0.001)。仅在IOPcc与GAT之间,CH与眼压测量差异显示出显著相关性(r = 0.830,P < 0.001)。

结论

NCT测得的平均眼压与GAT无显著差异,但ORA测得的眼压显著高于其他两种设备。