Auditory Implant Centre, Guy's and St. Thomas' NHS Trust, London, UK.
Clin Otolaryngol. 2010 Jun;35(3):190-7. doi: 10.1111/j.1749-4486.2010.02140.x.
To evaluate the hearing preservation rate and speech perception scores in patients with profound high frequency hearing loss and acoustically aidable low frequency hearing, managed with the MED-EL electric acoustic stimulation system referenced to the insertion depth of the electrode array.
Retrospective data analysis.
Ten patients implanted at the Auditory Implant Centre, Guy's and St Thomas's Hospital, London, UK.
Pure tone audiometry, speech perception tests and electrode insertion depth angle.
Postoperatively, functional hearing preservation allowing electric acoustic stimulation was achieved in eight patients and total preservation of residual hearing in five patients with follow-up periods of more than 12 months. Three of four (75%) patients with an insertion depth of >360 degrees had a threshold shift of >25 dB, and all four patients had a threshold shift of >10 dB. All patients with total hearing preservation had the electrode inserted up to 360 degrees at maximum. Overall, speech perception outcomes increased significantly and hearing impairment was significantly reduced after electric acoustic stimulation or electric stimulation alone as compared with the preoperative scores.
Electric acoustic stimulation provides significant benefit to individuals with profound high frequency hearing loss. Studies with larger number of patients are needed to establish the optimal electrode insertion angle as well as to further analyse the benefit of electric acoustic stimulation.
评估 MED-EL 电声刺激系统参照电极阵列插入深度管理的高频重度听力损失和可听低频听力患者的听力保存率和言语感知评分。
回顾性数据分析。
10 名在英国伦敦盖伊和圣托马斯医院听觉植入中心接受植入的患者。
纯音听阈测试、言语感知测试和电极插入深度角度。
术后 8 例患者获得允许电声刺激的功能性听力保存,5 例患者获得超过 12 个月的随访期内残余听力的完全保存。插入深度>360 度的 4 例中有 3 例(75%)阈值移位>25dB,4 例患者阈值移位>10dB。所有听力完全保存的患者电极插入深度最大可达 360 度。总体而言,与术前评分相比,电声刺激或单独电刺激后言语感知结果显著提高,听力障碍显著降低。
电声刺激为高频重度听力损失患者提供了显著的益处。需要进行更多患者的研究来确定最佳电极插入角度,并进一步分析电声刺激的益处。