RTVue OCT 与 GDx VCC 测量视网膜神经纤维层的对比研究。

Comparative study of retinal nerve fibre layer measurement by RTVue OCT and GDx VCC.

机构信息

Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing, People's Republic of China.

出版信息

Br J Ophthalmol. 2011 Apr;95(4):509-13. doi: 10.1136/bjo.2009.163493. Epub 2010 Jul 23.

Abstract

AIM

To evaluate the diagnostic performances and correlations of retinal nerve fibre layer (RNFL) thickness measured by RTVue OCT and GDx variable corneal compensation (VCC).

METHODS

The total and regional RNFL thickness were measured by RTVue OCT and GDx VCC in 62 normal eyes and 72 glaucomatous eyes of Chinese subjects. The RNFL thickness profiles of normal and glaucomatous eyes by RTVue OCT are plotted. Correlations of RNFL thickness measured by RTVue OCT and GDx VCC were assessed using the Pearson correlation. The discriminating abilities of the two techniques for detection of glaucoma were compared by the area under the receiver operating characteristic curves (AUC).

RESULTS

RTVue OCT demonstrated double hump patterns in the RNFL profiles. In both normal and glaucomatous subjects, the peaks were located in the superotemporal (ST) and inferotemporal (IT) regions, and the troughs were located at the nasal (NU+NL) and temporal (TU+TL) regions. Despite poor agreement, a high correlation (r=0.821) was found between the mean RNFL measurements by RTVue OCT and GDx VCC. For RTVue OCT, the highest AUCs were mean RNFL (AUC=0.914) and inferior mean RNFL (AUC=0.909). The nerve fibre indicator (AUC=0.856) and inferior RNFL (AUC=0.852) achieved the highest AUCs among all the GDx VCC measurements. The mean RNFL in RTVue OCT had the greatest AUC in the two devices. There was a significant difference in comparing the AUCs of the mean RNFL thickness obtained by RTVue OCT and GDx VCC (p=0.009).

CONCLUSIONS

Although there were absolute value differences in RNFL thickness, a high correlation was observed between RTVue OCT and GDx VCC. RTVue OCT shows a reasonable ability to distinguish normal from glaucomatous eyes.

摘要

目的

评估 RTVue OCT 与 GDx 可变角膜补偿(VCC)测量的视网膜神经纤维层(RNFL)厚度的诊断性能和相关性。

方法

对 62 例正常眼和 72 例中国青光眼患者的 RTVue OCT 和 GDx VCC 进行总 RNFL 和局部 RNFL 厚度测量。绘制 RTVue OCT 正常眼和青光眼眼的 RNFL 厚度图谱。采用 Pearson 相关评估 RTVue OCT 和 GDx VCC 测量的 RNFL 厚度之间的相关性。通过受试者工作特征曲线(AUC)下面积比较两种技术检测青光眼的能力。

结果

RTVue OCT 在 RNFL 图谱中显示出双峰模式。在正常和青光眼患者中,峰位于上方颞侧(ST)和下方颞侧(IT)区域,谷位于鼻侧(NU+NL)和颞侧(TU+TL)区域。尽管一致性差,但 RTVue OCT 平均 RNFL 测量值与 GDx VCC 之间存在高度相关性(r=0.821)。对于 RTVue OCT,平均 RNFL(AUC=0.914)和下方平均 RNFL(AUC=0.909)的 AUC 最高。所有 GDx VCC 测量中,神经纤维指标(AUC=0.856)和下方 RNFL(AUC=0.852)的 AUC 最高。RTVue OCT 中的平均 RNFL 在两种设备中的 AUC 最高。RTVue OCT 和 GDx VCC 获得的平均 RNFL 厚度 AUC 之间存在显著差异(p=0.009)。

结论

尽管 RNFL 厚度存在绝对值差异,但 RTVue OCT 与 GDx VCC 之间存在高度相关性。RTVue OCT 具有合理的区分正常眼与青光眼眼的能力。