Ophthalmology Department, University of São Paulo, Brazil.
J Refract Surg. 2011 Apr;27(4):287-92. doi: 10.3928/1081597X-20100714-01. Epub 2010 Jul 15.
To determine whether implantation of an intraocular lens (IOL) with an aspheric surface (Akreos AO, Bausch & Lomb Inc) results in reduced ocular aberrations (spherical aberration) and improved Strehl ratio and modulation transfer function (MTF) after cataract surgery.
In an intraindividual, randomized, double-masked, prospective study of 50 eyes (25 patients) with bilateral cataract, an IOL with modified anterior and posterior surfaces (Akreos AO) was implanted in one eye and a biconvex IOL with spherical surfaces (Akreos Fit, Bausch & Lomb Inc) implanted in the fellow eye. Ocular aberrations, Strehl ratio, and MTF curve with 4.5-, 5.0-, and 6.0-mm pupils were measured with a NIDEK OPD-Scan dynamic retinoscopy aberrometer 3 months after surgery. Uncorrected and corrected distance visual acuity (UDVA and CDVA, respectively) were also measured.
No statistically significant difference was noted between eyes in postoperative UDVA and CDVA at 1 month. At 3 months, the Akreos AO IOL group obtained statistically significant lower values of higher order and spherical aberrations with 4.5-, 5.0-, and 6.0-mm pupil diameters than the Akreos Fit IOL group (P<.05). The value of Strehl ratio was statistically significantly higher in eyes with the Akreos AO IOL for 4.5- and 6.0-mm pupils (P<.05). The MTF curve was better in the Akreos AO IOL group in 4.5-, 5.0-, and 6.0-mm pupils (P<.05).
The aspheric Akreos AO IOL induced significantly less spherical aberration than the Akreos Fit IOL for 4.5-, 5.0-, and 6.0-mm pupils. Modulation transfer function and Strehl ratio were also better in eyes implanted with the Akreos AO IOL than the Akreos Fit.
确定植入具有非球面表面(Akreos AO,Bausch & Lomb Inc)的人工晶状体(IOL)是否会减少白内障手术后的眼像差(球差)并提高斯特列尔比和调制传递函数(MTF)。
在一项针对 50 只眼(25 例患者)的双眼白内障的个体内、随机、双盲、前瞻性研究中,一只眼植入具有改良前后表面的 IOL(Akreos AO),另一只眼植入具有球面的双凸 IOL(Akreos Fit,Bausch & Lomb Inc)。术后 3 个月,使用 NIDEK OPD-Scan 动态视网膜验光仪测量眼像差、斯特列尔比和 4.5、5.0 和 6.0mm 瞳孔的 MTF 曲线。还测量了未矫正和矫正远视力(UDVA 和 CDVA,分别)。
术后 1 个月,双眼 UDVA 和 CDVA 无统计学差异。术后 3 个月,Akreos AO IOL 组与 Akreos Fit IOL 组相比,4.5、5.0 和 6.0mm 瞳孔直径的高阶像差和球差的测量值明显较低(P<.05)。Akreos AO IOL 组 4.5 和 6.0mm 瞳孔的斯特列尔比明显较高(P<.05)。Akreos AO IOL 组在 4.5、5.0 和 6.0mm 瞳孔时的 MTF 曲线更好(P<.05)。
与 Akreos Fit IOL 相比,Akreos AO 非球面 IOL 在 4.5、5.0 和 6.0mm 瞳孔时引起的球差明显更小。植入 Akreos AO IOL 的眼的调制传递函数和斯特列尔比也优于 Akreos Fit IOL。