How reliable is the Goldmann tonometer as a standard?

作者信息

Grolman B, Myers K J, Lalle P

出版信息

J Am Optom Assoc. 1990 Nov;61(11):857-62.

PMID:2081830
Abstract

In a clinical comparison of an XPERT NCT non-contact tonometer with Goldmann applanation tonometry (GAT), reported in the companion article, we concluded that the XPERT is a highly accurate and reliable instrument. By design, our evaluation differed significantly from earlier clinical studies in its use of multiple references (four calibrated GAT reference tonometers and seven GAT operators). Five clinicians were responsible for GAT measurement of 400 of a 620 eye population. Each clinician's population sample was essentially drawn at random from the common pool. XPERT vs. GAT performance indices were calculated for each of the five GAT operators. Because XPERT's measurements are essentially free of operator influence, it is reasonable to conclude that differences found among these five "XPERT performance" findings were primarily attributable to differences in GAT operator technique and endpoint judgment. Our findings disclose compelling evidence that GAT's accuracy and reliability depends on individual operator technique and experience. They indicate that in a clinical comparison, the protocol safeguards proposed by Wittenberg need be applied to ensure the integrity of studies using GAT as the reference tonometer.

摘要

在配套文章中报道的一项关于XPERT NCT非接触眼压计与Goldmann压平眼压计(GAT)的临床比较中,我们得出结论:XPERT是一种高度准确且可靠的仪器。从设计上看,我们的评估与早期临床研究有显著不同,我们使用了多个参考标准(四台校准过的GAT参考眼压计和七名GAT操作人员)。五名临床医生负责对620只眼中的400只眼进行GAT测量。每位临床医生的样本群体基本上是从共同的样本库中随机抽取的。针对五名GAT操作人员中的每一位,计算了XPERT与GAT的性能指标。由于XPERT的测量基本上不受操作人员的影响,因此可以合理地得出结论:在这五个“XPERT性能”结果中发现的差异主要归因于GAT操作人员技术和终点判断的差异。我们的研究结果揭示了令人信服的证据,表明GAT的准确性和可靠性取决于个体操作人员的技术和经验。这些结果表明,在临床比较中,需要应用Wittenberg提出的方案保障措施,以确保将GAT用作参考眼压计的研究的完整性。

相似文献

引用本文的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验