Patients with ocular hypertension (OHT) and glaucoma are increasingly reviewed in virtual clinics. As the clinician is not present during the patient's visit it is important that measurements obtained are reliable. The aim of this study was to examine agreement between Goldmann Applanation Tonometry (GAT) intraocular pressure (IOP) measurements (obtained by ophthalmologists and ophthalmic nurses) and a newer automated tonometer-the Ocular Response Analyzer (ORA).
A prospective study was conducted including 116 eyes of 116 patients with glaucoma and OHT. All subjects had GAT IOP obtained by a nurse and ophthalmologist and ORA IOP by a technician. The order of testing was randomised and previous measurements were masked. Agreement was examined using Bland-Altman plots and 95% limits of agreement (LoA). Intraclass correlation coefficients (ICC) of repeat GAT and ORA measurements were calculated.
Patients had a mean age of 70 ± 13 years. The 95% LoA between ophthalmologist (mean: 14.6 ± 4.3 mmHg) and nurse (mean: 15.0 ± 4.0 mmHg) GAT measurements was ±5.21 mmHg, whereas the 95% LoA between repeat ORA IOPg (mean: 13.8 ± 4.7 mmHg) was ±2.52 mmHg. There was no proportional bias. The ICC was 0.972 for repeat IOPg compared with 0.863 for repeat GAT.
There was only moderate agreement between GAT IOP measurements obtained by nurses working in the virtual clinic and ophthalmologists. Agreement between ORA IOP and ophthalmologists' GAT IOP was better and ORA produced more repeatable measurements, providing evidence it may be a more reliable tool for IOP assessment in virtual clinics.
患有眼高压(OHT)和青光眼的患者越来越多地在虚拟诊所中接受检查。由于在患者就诊期间临床医生不在场,因此获得的测量值必须可靠。本研究的目的是检查 Goldmann 压平眼压计(GAT)眼压(由眼科医生和眼科护士获得)与新型自动化眼压计-Ocular Response Analyzer(ORA)之间的一致性。
进行了一项前瞻性研究,共纳入 116 例青光眼和 OHT 患者的 116 只眼。所有受试者均由护士和眼科医生进行 GAT 眼压测量,由技术员进行 ORA 眼压测量。测试顺序是随机的,以前的测量值是屏蔽的。使用 Bland-Altman 图和 95%一致性界限(LoA)来检查一致性。计算了重复 GAT 和 ORA 测量的组内相关系数(ICC)。
患者的平均年龄为 70±13 岁。眼科医生(平均:14.6±4.3mmHg)和护士(平均:15.0±4.0mmHg)GAT 测量值之间的 95%LoA 为±5.21mmHg,而重复 ORA IOPg(平均:13.8±4.7mmHg)之间的 95%LoA 为±2.52mmHg。没有比例偏差。与重复 GAT 相比,IOPg 的 ICC 为 0.972,为 0.863。
在虚拟诊所中工作的护士获得的 GAT 眼压测量值与眼科医生之间仅存在中度一致性。ORA 眼压与眼科医生的 GAT 眼压之间的一致性更好,并且 ORA 产生了更可重复的测量值,这为其在虚拟诊所中进行眼压评估可能是一种更可靠的工具提供了证据。