Arterial oxygen status determined with routine pH/blood gas equipment and multi-wavelength hemoximetry: reference values, precision, and accuracy.

作者信息

Siggaard-Andersen O, Wimberley P D, Fogh-Andersen N, Gøthgen I H

机构信息

Department of Clinical Chemistry, Herlev Hospital, University of Copenhagen, Denmark.

出版信息

Scand J Clin Lab Invest Suppl. 1990;203:57-66. doi: 10.3109/00365519009087492.

Abstract

We measured pH, pCO2, pO2, oxygen saturation, total hemoglobin concentration, and fractions of carboxy- and methemoglobin in arterial blood samples from 35 healthy adults. We used a new algorithm to calculate active hemoglobin concentration, total oxygen concentration, actual half-saturation tension, 2,3-diphosphoglycerate concentration, estimated functional shunt, oxygen extraction tension px (for extracting 2.3 mmol of oxygen per liter of blood, values below 4.5 kPa indicating risk of tissue hypoxia), and the oxygen compensation factor Qx (the factor by which the cardiac output should rise to maintain a normal mixed venous pO2 of 5.0 kPa, factors above 1.5 indicating an extra burden on the heart). Analytical precision was evaluated by duplicate determinations. The accuracy of the half-saturation tension was evaluated by comparison with values for simultaneously drawn venous blood, the accuracy of the calculated concentration of 2,3-diphosphoglycerate by comparison with direct enzymatic measurements. We conclude that all the variables may be determined with sufficient accuracy and precision in healthy adults, provided the oxygen saturation is less than 0.97 and the measurements are performed according to the highest state of the art.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验