Department of Imaging Physics, The University of Texas M.D. Anderson Cancer Center, Houston, TX 77030, USA.
J Vasc Interv Radiol. 2010 Nov;21(11):1725-32. doi: 10.1016/j.jvir.2010.07.022.
To evaluate the accuracy of computer simulation in predicting the thermal damage region produced by a radiofrequency (RF) ablation procedure in an in vitro perfused bovine liver model. The thermal dose end point in the liver model is used to assess quantitatively computer prediction for use in prospective treatment planning of RF ablation procedures.
Geometric details of the tri-cooled tip electrode were modeled. The resistive heating of a pulsed voltage delivery was simulated in four dimensions using finite element models (FEM) implemented on high-performance parallel computing architectures. A range of physically realistic blood perfusion parameters, 3.6-53.6 kg/sec/m(3), was considered in the computer model. An Arrhenius damage model was used to predict the thermal dose. Dice similarity coefficients (DSC) were the metric of comparison between computational predictions and T1-weighted contrast-enhanced images of the damage obtained from a RF procedure performed on an in vitro perfused bovine liver model.
For a perfusion parameter greater than 16.3 kg/sec/m(3), simulations predict the temporal evolution of the damaged volume is perfusion limited and will reach a maximum value. Over a range of physically meaningful perfusion values, 16.3-33.1 kg/sec/m(3), the predicted thermal dose reaches the maximum damage volume within 2 minutes of the delivery and is in good agreement (DSC > 0.7) with experimental measurements obtained from the perfused liver model.
As measured by the computed volumetric DSC, computer prediction accuracy of the thermal dose shows good correlation with ablation lesions measured in vitro in perfused bovine liver models over a range of physically realistic perfusion values.
评估计算机模拟在预测射频(RF)消融程序在体外灌注牛肝模型中产生的热损伤区域的准确性。该肝模型中的热剂量终点用于评估计算机预测在 RF 消融程序的前瞻性治疗计划中的定量使用。
对三冷尖端电极的几何细节进行建模。使用在高性能并行计算架构上实现的有限元模型(FEM)对脉冲电压输送的电阻加热进行了四维模拟。在计算机模型中考虑了 3.6-53.6 kg/sec/m(3)的一系列物理上现实的血流灌注参数。使用阿仑尼乌斯损伤模型来预测热剂量。骰子相似系数(DSC)是比较计算机预测与通过在体外灌注牛肝模型上进行的 RF 程序获得的损伤的 T1 加权对比增强图像之间差异的度量。
对于灌注参数大于 16.3 kg/sec/m(3),模拟预测损伤体积的时间演化是灌注限制的,并将达到最大值。在 16.3-33.1 kg/sec/m(3)的一系列有意义的物理灌注值范围内,预测的热剂量在输送后 2 分钟内达到最大损伤体积,并且与从灌注肝模型获得的实验测量值吻合良好(DSC>0.7)。
根据计算的体积 DSC 测量,在物理上真实的灌注值范围内,计算机对热剂量的预测准确性与体外灌注牛肝模型中的消融病变具有良好的相关性。