Koessler Laurent, Cecchin Thierry, Ternisien Eric, Maillard Louis
Centre de Recherche en Automatique de Nancy (CRAN), Nancy-Université, CNRS and Centre Hospitalier Universitaire de Nancy, France.
Annu Int Conf IEEE Eng Med Biol Soc. 2010;2010:3707-10. doi: 10.1109/IEMBS.2010.5627659.
This paper describes and assesses for the first time the use of a handheld 3D laser scanner for scalp EEG sensor localization and co-registration with magnetic resonance images. Study on five subjects showed that the scanner had an equivalent accuracy, a better repeatability, and was faster than the reference electromagnetic digitizer. According to electrical source imaging, somatosensory evoked potentials experiments validated its ability to give precise sensor localization. With our automatic labeling method, the data provided by the scanner could be directly introduced in the source localization studies.
本文首次描述并评估了手持式3D激光扫描仪用于头皮脑电图传感器定位以及与磁共振图像进行配准的情况。对五名受试者的研究表明,该扫描仪具有同等的准确性、更好的可重复性,并且比参考电磁数字化仪速度更快。根据电源成像,体感诱发电位实验验证了其进行精确传感器定位的能力。采用我们的自动标记方法,扫描仪提供的数据可直接引入源定位研究中。