Radiotherapy Physics, Belfast Health and Social Care Trust, Northern Ireland Cancer Centre, Belfast City Hospital Site, Lisburn Road, Belfast BT9 7AB, UK.
Phys Med Biol. 2011 Jan 7;56(1):N11-9. doi: 10.1088/0031-9155/56/1/N02. Epub 2010 Nov 30.
Analysis of Varian linear accelerator output trends is reported. Two groups, consisting of four matched Varian 2100C/D and four matched Varian 600C/D accelerators, with different designs of monitor chamber, have been investigated and the data acquired from regular calibrated ion chamber/electrometer measurements of the output performance of the eight accelerators analysed. The trend of machine output with time, having removed the effect of adjusting the monitor chamber response, was compared on a monthly and annual basis for monitor chambers with ages ranging between 1 year and 7 years. The results indicate that the response is generally consistent within each set of accelerators with different monitor chamber designs. Those used in a Varian 600C/D machine result in a reduction in measured output over time, with an average monthly reduction of 0.35 ± 0.09% over the course of the first 4 years of use. The chambers used in a 2100C/D accelerator result in an increase in measured output over time, with an average monthly increase of 0.26 ± 0.09% over the course of the first 4 years of use. The output increase then reduces towards the end of this period of time, with the average monthly change falling to -0.03 ± 0.02% for the following 3 years. The output response trend was similar for all clinical energies used on the 2100C/D accelerators--6, 15 MV x-ray beams, and 4, 6, 9, 12, 16 and 20 MeV electron beams. By tracking these changes it has been possible to predict the response over time to allow appropriate adjustments in monitor chamber response to maintain a measured accelerator output within tolerance and give confidence in performance. It has also provided data to indicate the need for planned preventative intervention and indicate if the monitor chamber response is behaving as expected.
报道了瓦里安直线加速器输出趋势的分析。研究了两组共 8 台加速器,每组包括 4 台匹配的瓦里安 2100C/D 和 4 台匹配的瓦里安 600C/D 加速器,它们的监测腔设计不同,并从对 8 台加速器输出性能的定期校准离子室/静电计测量中获取数据。在去除了监测腔响应调整的影响后,按月和按年比较了具有 1 至 7 年不同年龄的监测腔的机器输出趋势。结果表明,不同监测腔设计的加速器中,响应通常是一致的。在瓦里安 600C/D 机器中使用的监测腔会导致测量输出随时间减少,在使用的前 4 年中,每月平均减少 0.35 ± 0.09%。在 2100C/D 加速器中使用的监测腔会导致测量输出随时间增加,在使用的前 4 年中,每月平均增加 0.26 ± 0.09%。然后,在这段时间结束时,输出增加量减少,在接下来的 3 年中,每月平均变化率降至-0.03 ± 0.02%。2100C/D 加速器上使用的所有临床能量(6、15 MV X 射线束以及 4、6、9、12、16 和 20 MeV 电子束)的输出响应趋势相似。通过跟踪这些变化,可以预测随时间的响应,以在监测腔响应中进行适当调整,从而保持测量加速器输出在容差范围内,并对性能充满信心。它还提供了数据,以指示需要计划的预防性干预,并指示监测腔响应是否按预期运行。