Suppr超能文献

Acute complication due to impella 2.5 device (superficial femoral artery thrombosis): managed successfully with novel aspiration thrombectomy catheter (pronto v3).

作者信息

Bhat Tariq M, Waked Alain, Teli Sumaya, Lafferty James, Gala Bhavesh

机构信息

Department of Medicine, Staten Island University Hospital, New York, USA.

出版信息

Clin Med Insights Cardiol. 2011 Feb 10;5:17-21. doi: 10.4137/CMC.S6157.

Abstract

The Impella recover LP 2.5 is a percutaneous left ventricular assist device (LVAD) recently approved for use in patients undergoing high risk percutaneous coronary intervention (PCI) and also in cases of cardiogenic shock. There is limited evidence available in literature about its safety, especially with regards to the incidence of local vascular complications, their management and long-term implications. We report here the first case of a serious local vascular complication-superficial femoral artery thrombus formation during Impella recover LP 2.5 use in a high risk PCI which was managed successfully with novel aspiration thrombectomy catheter (Pronto V3), which in itself is the first reported use of Pronto V3 in such a vascular complication.

摘要
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e816/3041237/78d3465a6229/cmc-2011-017f1.jpg