Comparison of automated analysis of Cirrus HD OCT spectral-domain optical coherence tomography with stereo photographs of the optic disc.

机构信息

Department of Ophthalmology, University of Miami Miller School of Medicine, Miami, FL, USA.

出版信息

Ophthalmology. 2011 Jul;118(7):1348-57. doi: 10.1016/j.ophtha.2010.12.008. Epub 2011 Mar 12.

Abstract

OBJECTIVE

To evaluate a new automated analysis of optic disc images obtained by spectral-domain optical coherence tomography (SD OCT). Areas of the optic disc, cup, and neural rim in SD OCT images were compared with these areas from stereoscopic photographs to represent the current traditional optic nerve evaluation. The repeatability of measurements by each method was determined and compared.

DESIGN

Evaluation of diagnostic technology.

PARTICIPANTS

One hundred nineteen healthy eyes, 23 eyes with glaucoma, and 7 glaucoma suspect eyes.

METHODS

Optic disc and cup margins were traced from stereoscopic photographs by 3 individuals independently. Optic disc margins and rim widths were determined automatically in SD OCT. A subset of photographs was examined and traced a second time, and duplicate SD OCT images also were analyzed.

MAIN OUTCOME MEASURES

Agreement among photograph readers, between duplicate readings, and between SD OCT and photographs were quantified by the intraclass correlation coefficient (ICC), by the root mean square, and by the standard deviation of the differences.

RESULTS

Optic disc areas tended to be slightly larger when judged in photographs than by SD OCT, whereas cup areas were similar. Cup and optic disc areas showed good correlation (0.8) between the average photographic reading and SD OCT, but only fair correlation of rim areas (0.4). The SD OCT was highly reproducible (ICC, 0.96-0.99). Each reader also was consistent with himself on duplicate readings of 21 photographs (ICC, 0.80-0.88 for rim area and 0.95-0.98 for all other measurements), but reproducibility was not as good as SD OCT. Measurements derived from SD OCT did not differ from photographic readings more than the readings of photographs by different readers differed from each other.

CONCLUSIONS

Designation of the cup and optic disc boundaries by an automated analysis of SD OCT was within the range of variable designations by different readers from color stereoscopic photographs, but use of different landmarks typically made the designation of the optic disc size somewhat smaller in the automated analysis. There was better repeatability among measurements from SD OCT than from among readers of photographs. The repeatability of automated measurement of SD OCT images is promising for use both in diagnosis and in monitoring of progression.

摘要

目的

评估一种新的应用于频域光学相干断层扫描(SD OCT)的视神经盘图像自动分析方法。通过比较 SD OCT 图像中视神经盘、杯和神经边缘的区域与立体照片中的这些区域,来代表当前传统的视神经评估。分别确定并比较了这两种方法的测量重复性。

设计

诊断技术评估。

参与者

119 只健康眼,23 只青光眼眼和 7 只青光眼疑似眼。

方法

由 3 名独立人员分别从立体照片中追踪视神经盘和杯缘。SD OCT 自动确定视神经盘边缘和边缘宽度。检查了照片的子集并进行了第二次追踪,并且还对重复的 SD OCT 图像进行了分析。

主要观察指标

通过组内相关系数(ICC)、均方根和差值标准差,量化照片阅读者之间、两次阅读之间以及 SD OCT 与照片之间的一致性。

结果

与 SD OCT 相比,通过照片判断的视神经盘区域往往稍大,而杯区相似。杯和视神经盘区域在平均照片读数和 SD OCT 之间具有良好的相关性(0.8),但边缘区域的相关性仅为中等(0.4)。SD OCT 具有高度的可重复性(ICC,0.96-0.99)。每位阅读者对 21 张照片的两次重复阅读也具有一致性(边缘区域的 ICC 为 0.80-0.88,所有其他测量值的 ICC 为 0.95-0.98),但可重复性不如 SD OCT 好。SD OCT 得出的测量值与不同读者的照片读数之间的差异,并不比不同读者的照片读数之间的差异更大。

结论

SD OCT 自动分析指定杯和视神经盘边界的范围与不同读者从彩色立体照片中指定边界的范围相似,但使用不同的标记通常会使自动分析中视神经盘大小的指定稍小。SD OCT 之间的测量重复性优于照片阅读者之间的重复性。SD OCT 图像自动测量的重复性对于诊断和监测进展都具有很大的应用前景。