Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, CH-1211 Geneva, Switzerland.
Phys Med Biol. 2011 May 21;56(10):3091-106. doi: 10.1088/0031-9155/56/10/013. Epub 2011 Apr 20.
The Ingenuity TF PET-MRI is a newly released whole-body hybrid PET-MR imaging system with a Philips time-of-flight GEMINI TF PET and Achieva 3T X-series MRI system. Compared to PET-CT, modifications to the positron emission tomography (PET) gantry were made to avoid mutual system interference and deliver uncompromising performance which is equivalent to the standalone systems. The PET gantry was redesigned to introduce magnetic shielding for the photomultiplier tubes (PMTs). Stringent electromagnetic noise requirements of the MR system necessitated the removal of PET gantry electronics to be housed in the PET-MR equipment room. We report the standard NEMA measurements for the PET scanner. PET imaging and performance measurements were done at Geneva University Hospital as described in the NEMA Standards NU 2-2007 manual. The scatter fraction (SF) and noise equivalent count rate (NECR) measurements with the NEMA cylinder (20 cm diameter) were repeated for two larger cylinders (27 cm and 35 cm diameter), which better represent average and heavy patients. A NEMA/IEC torso phantom was used for overall assessment of image quality. The transverse and axial resolution near the center was 4.7 mm. Timing and energy resolution of the PET-MR system were measured to be 525 ps and 12%, respectively. The results were comparable to PET-CT systems demonstrating that the effect of design modifications required on the PET system to remove the harmful effect of the magnetic field on the PMTs was negligible. The absolute sensitivity of this scanner was 7.0 cps kBq(-1), whereas SF was 26%. NECR measurements performed with cylinders having three different diameters, and image quality measurements performed with IEC phantom yielded excellent results. The Ingenuity TF PET-MRI represents the first commercial whole-body hybrid PET-MRI system. The performance of the PET subsystem was comparable to the GEMINI TF PET-CT system using phantom and patient studies. It is conceived that advantages of hybrid PET-MRI will become more evident in the near future.
Ingenuity TF PET-MRI 是一款全新发布的全身混合式 PET-MRI 成像系统,配备飞利浦飞行时间 GEMINI TF PET 和 Achieva 3T X 系列 MRI 系统。与 PET-CT 相比,对正电子发射断层扫描(PET)龙门架进行了修改,以避免相互系统干扰,并提供与独立系统相当的卓越性能。PET 龙门架经过重新设计,引入了光电倍增管(PMT)的磁屏蔽。磁共振系统严格的电磁噪声要求需要将 PET 龙门架电子设备移至 PET-MR 设备室中。我们报告了 PET 扫描仪的标准 NEMA 测量结果。根据 NEMA 标准 NU 2-2007 手册的描述,在日内瓦大学医院进行了 PET 成像和性能测量。使用 NEMA 圆柱(20cm 直径)重复进行了散射分数(SF)和噪声等效计数率(NECR)测量,还使用了两个更大的圆柱(27cm 和 35cm 直径)进行了重复测量,这更好地代表了平均体型和超重体型的患者。使用 NEMA/IEC 躯干体模对整体图像质量进行了评估。在中心附近,横向和轴向分辨率为 4.7mm。测量 PET-MR 系统的时间和能量分辨率分别为 525ps 和 12%。结果与 PET-CT 系统相当,表明为消除磁场对 PMT 的有害影响而对 PET 系统进行的设计修改的效果可以忽略不计。该扫描仪的绝对灵敏度为 7.0cpskBq(-1),而 SF 为 26%。使用三个不同直径的圆柱进行的 NECR 测量以及使用 IEC 体模进行的图像质量测量均取得了优异的结果。Ingenuity TF PET-MRI 是首款商用的全身混合式 PET-MRI 系统。使用体模和患者研究进行的 PET 子系统性能与 GEMINI TF PET-CT 系统相当。可以预见,混合式 PET-MRI 的优势在不久的将来会变得更加明显。