Glaucoma Center at the Montchoisi Clinic, Lausanne, Switzerland.
J Glaucoma. 2012 Oct-Nov;21(8):539-44. doi: 10.1097/IJG.0b013e31821dac43.
To evaluate the tolerability, comfort, and reliability of the signal transmission of an ocular Sensor used for 24-hour intraocular pressure fluctuation monitoring in humans.
In this uncontrolled open trial involving 10 healthy volunteers, an 8.7-mm radius prototype ocular telemetry Sensor (SENSIMED Triggerfish, Lausanne, Switzerland) and an orbital bandage containing a loop antenna were applied and connected to a portable recorder after full eye examination. Best-corrected visual acuity and position, surface wetting ability, and mobility of the Sensor were assessed after 5 and 30 minutes, 4, 12, and 24 hours. Subjective wearing comfort was scored and activities documented in a logbook. After Sensor removal, a full eye examination was repeated and the recorded signal analyzed.
The comfort score was high and did not fluctuate significantly over time. The mobility of the Sensor was limited across follow-up visits and its surface wetting ability remained good. Best-corrected visual acuity was significantly reduced during Sensor wear and immediately after its removal (from 1.07 before, to 0.85 after, P value 0.008). Three subjects developed a mild, transient corneal abrasion. In all but 1 participant, we obtained usable data of a telemetric signal recording with sufficient sensitivity to depict ocular pulsation.
This 24-hour trial has encouraging results on the tolerability and functionality of the ocular telemetric Sensor for intraocular pressure fluctuation monitoring. Further studies with different Sensor radii conducted on a larger study population are needed to improve comfort, precision, and interpretation of the telemetric signal.
评估用于监测人类 24 小时眼压波动的眼部传感器的信号传输的耐受性、舒适性和可靠性。
在这项涉及 10 名健康志愿者的非对照开放性试验中,应用了一个 8.7mm 半径原型眼部遥测传感器(瑞士洛桑 Sensimed Triggerfish)和一个包含环形天线的眼眶绷带,在全面眼部检查后,将其连接到便携式记录器。在 5 分钟和 30 分钟、4 小时、12 小时和 24 小时后,评估最佳矫正视力和位置、表面湿润能力以及传感器的活动性。主观佩戴舒适度评分,并在日志中记录活动情况。传感器移除后,重复全面眼部检查并分析记录的信号。
舒适度评分较高,且随时间无明显波动。传感器的活动性在随访期间受到限制,其表面湿润能力仍然良好。在佩戴传感器期间和移除后,最佳矫正视力显著降低(从佩戴前的 1.07 降至 0.85,P 值 0.008)。有 3 名受试者出现轻度、短暂的角膜擦伤。除 1 名参与者外,我们均获得了具有足够敏感性的遥测信号记录的可用数据,可用于描绘眼部搏动。
这项 24 小时试验对眼部遥测传感器用于眼压波动监测的耐受性和功能提供了令人鼓舞的结果。需要在更大的研究人群中进行不同传感器半径的进一步研究,以提高舒适度、精度和遥测信号的解读。