Transient changes in refractive error and corneal tomography after 24-h continuous monitoring of intraocular pressure patterns with a contact lens sensor.

作者信息

Miki Atsuya, Kumoi Miho, Maeda Naoyuki, Koh Shizuka, Matsushita Kenji, Nishida Kohji

机构信息

Department of Ophthalmology, Osaka University Graduate School of Medicine, 2-2 Yamada-Oka, Suita, Osaka, 565-0871, Japan.

Department of Ophthalmology, National Hospital Organization Osaka National Hospital, Osaka, Japan.

出版信息

Jpn J Ophthalmol. 2020 Mar;64(2):127-133. doi: 10.1007/s10384-020-00723-6. Epub 2020 Feb 13.

Abstract

PURPOSE

To evaluate changes in refractive error and corneal tomographic measurements after 24-h monitoring of intraocular pressure patterns with a contact lens sensor (CLS).

STUDY DESIGN

Prospective, longitudinal, observational study.

PATIENTS AND METHODS

Fourteen eyes of 14 consecutive patients who underwent 24-h CLS monitoring were enrolled. The objective spherical equivalent (SE) refractive error was measured using automated refraction and keratometry. The axial power, instantaneous power, and corneal thickness at the central, paracentral, midperipheral, and peripheral cornea were measured with swept-source anterior-segment optical coherence tomography. Measurements were performed at baseline, immediately after monitoring, and at follow-up visits 2-4 days after monitoring.

RESULTS

The myopic SE increased significantly (P < 0.001) from - 5.1 ± 4.2 to - 6.0 ± 4.0, D after 24-h monitoring with a CLS and returned to the premonitoring level at the second visit (- 5.3 ± 4.4 D, P = 0.315). Decreases in the midperipheral and peripheral axial powers and the paracentral instantaneous power and increases in the central instantaneous power and peripheral corneal thickness occurred after CLS monitoring.

CONCLUSION

Twenty-four-hour monitoring with a CLS resulted in significant increases in the myopic refractive error and corneal central steepening and midperipheral flattening. Although these changes are transient and do not negate the clinical merits of the CLS, clinicians should be aware of these potential adverse events. Three-dimensional evaluation of corneal deformation with anterior segment optical coherence tomography is a powerful tool for assessing and improving the safety, tolerability, and accuracy of CLS devices.

摘要

目的

使用隐形眼镜传感器(CLS)对眼压模式进行24小时监测后,评估屈光不正和角膜断层扫描测量值的变化。

研究设计

前瞻性、纵向、观察性研究。

患者和方法

纳入连续14例接受24小时CLS监测患者的14只眼。使用自动验光和角膜曲率测量法测量客观球镜等效(SE)屈光不正。使用扫频源眼前节光学相干断层扫描测量中央、旁中央、中周边和周边角膜的轴向屈光力、瞬时屈光力和角膜厚度。在基线、监测结束后立即以及监测后2 - 4天的随访时进行测量。

结果

使用CLS进行24小时监测后,近视SE从 - 5.1±4.2显著增加(P < 0.001)至 - 6.0±4.0 D,并在第二次随访时恢复到监测前水平( - 5.3±4.4 D,P = 0.315)。CLS监测后,中周边和周边轴向屈光力以及旁中央瞬时屈光力降低,中央瞬时屈光力和周边角膜厚度增加。

结论

使用CLS进行24小时监测导致近视屈光不正显著增加,角膜中央变陡和中周边变平。尽管这些变化是短暂的,且不否定CLS的临床优点,但临床医生应意识到这些潜在不良事件。使用眼前节光学相干断层扫描对角膜变形进行三维评估是评估和提高CLS设备安全性、耐受性和准确性的有力工具。