University Eye Hospital Wuerzburg, Wuerzburg, Germany.
Cornea. 2011 Oct;30(10):1078-82. doi: 10.1097/ICO.0b013e31820cd3d6.
We performed a comparative study using Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT) to measure intraocular pressure (IOP) in eyes with keratoconus.
IOP was measured in 114 eyes of 75 patients with keratoconus (51 men and 24 women; mean age, 36.1 ± 11.2 years) using GAT and DCT in randomized order. Central corneal thickness (CCT), minimal corneal thickness (MCT), and corneal topography were recorded using the Pentacam. Four groups according to Amsler's keratoconus classification were composed and analyzed for significant differences of CCT, MCT, GAT, and DCT results.
Mean CCT in the 114 keratoconus eyes was 481.1 ± 46.2 μm (range, 334-601 μm). Mean MCT was 453.3 ± 56.3 μm (range, 239-573 μm). Mean IOP measured using GAT was 13.1 ± 2.9 mm Hg, whereas mean IOP measured using DCT was 14.8 ± 2.6 mm Hg. Neither the results for GAT nor those for DCT showed a significant correlation with CCT (Pearson correlation: P < 0.05). Multifactorial analysis revealed that CCT and MCT, but not GAT and DCT, results were significantly different in corneas of varying curvatures.
This study shows that DCT measures IOP higher than GAT in eyes with keratoconus. In keratoconus, both methods seem to be independent of CCT and therefore are equally, but not interchangeably, applicable when monitoring IOP. Further analysis revealed that CCT and MCT are significantly different in corneas of varying Amsler grade.
我们使用 Goldmann 压平眼压计(GAT)和动态轮廓眼压计(DCT)对圆锥角膜患者的眼压(IOP)进行了比较研究。
我们以随机顺序使用 GAT 和 DCT 测量了 75 例(51 名男性和 24 名女性;平均年龄 36.1 ± 11.2 岁)圆锥角膜患者的 114 只眼的 IOP。使用 Pentacam 记录中央角膜厚度(CCT)、最小角膜厚度(MCT)和角膜地形图。根据 Amsler 圆锥角膜分类法将 CCT、MCT、GAT 和 DCT 的结果分为四组进行分析,以确定差异是否有统计学意义。
114 只圆锥角膜眼中的平均 CCT 为 481.1 ± 46.2μm(范围 334-601μm)。平均 MCT 为 453.3 ± 56.3μm(范围 239-573μm)。GAT 测量的平均 IOP 为 13.1 ± 2.9mmHg,而 DCT 测量的平均 IOP 为 14.8 ± 2.6mmHg。GAT 和 DCT 的结果均与 CCT 无显著相关性(Pearson 相关:P<0.05)。多因素分析显示,角膜曲率不同时,CCT 和 MCT 但不是 GAT 和 DCT 的结果有显著差异。
本研究表明,在圆锥角膜患者中,DCT 测量的眼压高于 GAT。在圆锥角膜中,这两种方法似乎都与 CCT 无关,因此在监测 IOP 时同样适用,但不能互换。进一步分析显示,角膜曲率不同时,CCT 和 MCT 有显著差异。