在植入MyoRing后圆锥角膜患者中使用Goldmann压平眼压计、ICare回弹眼压计和Tono-Pen测量眼压的比较。

Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation.

作者信息

Rateb Mahmoud, Abdel-Radi Mahmoud, Eldaly Zeiad, Elmohamady Mohamed Nagy, Noor El Din Asaad

机构信息

Department of Ophthalmology, Faculty of Medicine, Assiut University, Assiut, Egypt.

Tiba Eye Center, Assiut, Egypt.

出版信息

J Ophthalmol. 2019 May 9;2019:1964107. doi: 10.1155/2019/1964107. eCollection 2019.

Abstract

PURPOSE

To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers.

SETTING

Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016.

METHODS

Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by -test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland-Altman analysis was used to assess intertonometer agreement.

RESULTS

Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare ( value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen ( value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was -0.39 ± 2.59 mmHg. There was no significant correlation between the difference in IOP readings among any pair of devices and CCC or TCL. The Bland-Altman analysis showed a reasonable agreement between any pair of tonometers.

摘要

目的

评估在圆锥角膜患者植入MyoRing后,使用Goldmann压平眼压计(GAT)、ICare回弹式眼压计和Tono-Pen测量眼压的差异。评估中央角膜厚度(CCT)和最薄角膜位置(TCL)对不同眼压计测量眼压的影响。

设置

2015年2月至2016年11月在埃及的两个私人中心进行了前瞻性观察研究。

方法

招募了10例圆锥角膜患者的17只眼,这些患者均接受了MyoRing植入术。所有受试者均按随机顺序接受GAT、ICare和Tono-Pen眼压测量。使用Pentacam评估中央角膜厚度和最薄角膜位置。通过t检验评估眼压读数的平均差异。通过Pearson相关系数评估每对设备之间的相关性。采用Bland-Altman分析评估眼压计之间的一致性。

结果

评估了17只眼(10例患者)。GAT组、ICare组和Tono-Pen组的平均眼压读数分别为13.9±3.68、12.41±2.87和14.29±1.31mmHg。GAT/ICare与Tono-Pen/ICare的眼压读数之间存在显著差异(t值分别为0.032和0.002),而GAT/Tono-Pen之间无显著差异(t值为0.554)。GAT/ICare之间眼压测量的平均差异为1.49±2.6 mmHg,Tono-Pen/ICare为1.89±2.15 mmHg,GAT/Tono-Pen为-0.39±2.59 mmHg。任何一对设备之间的眼压读数差异与CCT或TCL均无显著相关性。Bland-Altman分析显示任何一对眼压计之间具有合理的一致性。