Comparison of in vitro hydrolysis, subcutaneous and intramedullary implantation to evaluate the strength retention of absorbable osteosynthesis implants.

作者信息

Vasenius J, Vainionpää S, Vihtonen K, Mäkelä A, Rokkanen P, Mero M, Törmälä P

机构信息

Department of Orthopaedics and Traumatology, University Central Hospital, Helsinki, Finland.

出版信息

Biomaterials. 1990 Sep;11(7):501-4. doi: 10.1016/0142-9612(90)90065-x.

Abstract

To evaluate the behaviour of mechanical properties of absorbable osteosynthesis implants in vivo, the strength retention of self-reinforced polyglycolide rods in distilled water at 37 degrees C in the subcutis and femoral medullary cavity of the rabbit was investigated. The self-reinforced polyglycolide rods lost their strength significantly faster in vivo than in vitro. The strength loss of the self-reinforced polyglycolide rods was only slightly faster in the medullary cavity than in the subcutis. As the removal of the implants from the medullary cavity became difficult 5-6 wk after implantation, it is suggested that subcutaneous implantation would be a suitable method to evaluate the strength retention of absorbable osteosynthesis implants.

摘要

为评估可吸收接骨植入物在体内的力学性能表现,研究了自制增强聚乙交酯棒在37℃蒸馏水中、兔皮下及股骨髓腔内的强度保持情况。自制增强聚乙交酯棒在体内的强度丧失明显快于体外。自制增强聚乙交酯棒在髓腔内的强度丧失仅比皮下稍快。由于植入后5-6周从髓腔取出植入物变得困难,建议皮下植入是评估可吸收接骨植入物强度保持的合适方法。