Int J Cardiovasc Imaging. 2012 Jun;28(5):1141-6. doi: 10.1007/s10554-011-9929-8. Epub 2011 Jul 31.
Increased aortic pulse wave velocity (PWV) results from loss of arterial compliance and is associated with unfavorable outcomes. Applanation tonometry (AT) is the most frequently applied method to assess PWV and deduce aortic compliance. The goal of this study was to compare the reproducibility of PWV measurements obtained with: (1) cross-correlation analysis of phase contrast magnetic resonance (PCMR) velocity data, and (2) applanation tonometry (AT). PWV was measured twice with each modality in 13 normal young volunteers (controls) and 9 older patients who had undergone a CT exam to evaluate coronary artery calcium. The coefficient of variation (CoV) between measurements was computed for each modality. There was no significant difference in PWV values obtained with AT and PCMR in controls or patients. The inter-scan reproducibility of PCMR was superior to AT in the controls (CoV: 3.4 ± 2.3% vs. 6.3 ± 4.0%, P = 0.03) but not in the older patients (7.4 ± 8.0% vs. 9.9 ± 9.6%, P = 0.32). PWV values were higher in patients than controls (5.6 ± 1.2 vs. 9.7 ± 2.8, P = 0.002). PCMR and AT yielded similar values for PWV in patients and volunteers. PCMR showed a superior reproducibility in young subjects but not in older patients.
脉搏波速度(PWV)的增加源于动脉顺应性的丧失,与不良预后相关。平板压力测量法(AT)是评估 PWV 和推断动脉顺应性最常用的方法。本研究的目的是比较两种方法测量 PWV 的重复性:(1)相位对比磁共振(PCMR)速度数据的互相关分析,和(2)平板压力测量法(AT)。对 13 名正常年轻志愿者(对照组)和 9 名接受 CT 检查评估冠状动脉钙的老年患者,分别用两种方法对 PWV 进行两次测量。计算每种方法测量的变异系数(CoV)。在对照组和患者中,AT 和 PCMR 测量的 PWV 值没有显著差异。在对照组中,PCMR 的扫描间可重复性优于 AT(CoV:3.4 ± 2.3%比 6.3 ± 4.0%,P = 0.03),但在老年患者中无差异(7.4 ± 8.0%比 9.9 ± 9.6%,P = 0.32)。患者的 PWV 值高于对照组(5.6 ± 1.2 比 9.7 ± 2.8,P = 0.002)。PCMR 和 AT 在患者和志愿者中均能得到相似的 PWV 值。在年轻受试者中,PCMR 具有更好的可重复性,但在老年患者中并非如此。