Department of Radiation Physics, UT MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030, USA.
Med Phys. 2011 Jul;38(7):3955-62. doi: 10.1118/1.3596533.
Optically stimulated luminescent detectors (OSLDs) are quickly gaining popularity as passive dosimeters, with applications in medicine for linac output calibration verification, brachytherapy source verification, treatment plan quality assurance, and clinical dose measurements. With such wide applications, these dosimeters must be characterized for numerous factors affecting their response. The most abundant commercial OSLD is the InLight/OSL system from Landauer, Inc. The purpose of this study was to examine the angular dependence of the nanoDot dosimeter, which is part of the InLight system.
Relative dosimeter response data were taken at several angles in 6 and 18 MV photon beams, as well as a clinical proton beam. These measurements were done within a phantom at a depth beyond the build-up region. To verify the observed angular dependence, additional measurements were conducted as well as Monte Carlo simulations in MCNPX.
When irradiated with the incident photon beams parallel to the plane of the dosimeter, the nanoDot response was 4% lower at 6 MV and 3% lower at 18 MV than the response when irradiated with the incident beam normal to the plane of the dosimeter. Monte Carlo simulations at 6 MV showed similar results to the experimental values. Examination of the results in Monte Carlo suggests the cause as partial volume irradiation. In a clinical proton beam, no angular dependence was found.
A nontrivial angular response of this OSLD was observed in photon beams. This factor may need to be accounted for when evaluating doses from photon beams incident from a variety of directions.
光激励发光探测器(OSLD)作为被动剂量计正在迅速普及,在医学中有广泛的应用,包括线性加速器输出校准验证、近距离放射治疗源验证、治疗计划质量保证和临床剂量测量。由于这些剂量计的应用如此广泛,因此必须对影响其响应的众多因素进行特性描述。最丰富的商业 OSLD 是 Landauer 公司的 InLight/OSL 系统。本研究的目的是研究纳米点剂量计的角度依赖性,它是 InLight 系统的一部分。
在 6MV 和 18MV 光子束以及临床质子束中,在体模中超过建成区深度的位置,在几个角度上获取相对剂量计响应数据。为了验证观察到的角度依赖性,进行了额外的测量以及在 MCNPX 中的蒙特卡罗模拟。
当用与探测器平面平行的入射光子束照射时,纳米点的响应比用垂直于探测器平面的入射光束照射时,在 6MV 时降低了 4%,在 18MV 时降低了 3%。6MV 的蒙特卡罗模拟显示出与实验值相似的结果。在蒙特卡罗模拟中对结果的检查表明,原因是部分体积照射。在临床质子束中,没有发现角度依赖性。
在光子束中观察到这种 OSLD 的非平凡角度响应。在评估来自各种方向的光子束照射的剂量时,可能需要考虑这个因素。