The purpose of this prospective multicenter study was to assess the safety and technical feasibility of volumetric Magnetic Resonance-guided High Intensity Focused Ultrasound (MR-HIFU) ablation for treatment of patients with symptomatic uterine fibroids.
Thirty-three patients with 36 fibroids were treated with volumetric MR-HIFU ablation. Treatment capability and technical feasibility were assessed by comparison of the Non-Perfused Volumes (NPVs) with MR thermal dose predicted treatment volumes. Safety was determined by evaluation of complications or adverse events and unintended lesions. Secondary endpoints were pain and discomfort scores, recovery time and length of hospital stay.
The mean NPV calculated as a percentage of the total fibroid volume was 21.7%. Correlation between the predicted treatment volumes and NPVs was found to be very strong, with a correlation coefficient r of 0.87. All patients tolerated the treatment well and were treated on an outpatient basis. No serious adverse events were reported and recovery time to normal activities was 2.3 ± 1.8 days.
This prospective multicenter study proved that volumetric MR-HIFU is safe and technically feasible for the treatment of symptomatic uterine fibroids.
• Magnetic-resonance-guided high intensity focused ultrasound allows non-invasive treatment of uterine fibroids. • Volumetric feedback ablation is a novel technology that allows larger treatment volumes • MR-guided ultrasound ablation of uterine fibroids appears safe using volumetric feedback.
本前瞻性多中心研究旨在评估容积磁共振引导高强度聚焦超声(MR-HIFU)消融治疗有症状子宫肌瘤患者的安全性和技术可行性。
33 名 36 个肌瘤患者接受了容积 MR-HIFU 消融治疗。通过比较非灌注体积(NPV)与 MR 热剂量预测治疗体积,评估治疗能力和技术可行性。通过评估并发症或不良事件和非预期病变来确定安全性。次要终点是疼痛和不适评分、恢复时间和住院时间。
计算的总肌瘤体积百分比的平均 NPV 为 21.7%。预测治疗体积与 NPV 之间存在很强的相关性,相关系数 r 为 0.87。所有患者均能很好地耐受治疗,并在门诊接受治疗。未报告严重不良事件,正常活动的恢复时间为 2.3±1.8 天。
本前瞻性多中心研究证明了容积 MR-HIFU 治疗有症状子宫肌瘤的安全性和技术可行性。
•磁共振引导高强度聚焦超声可实现子宫肌瘤的非侵入性治疗。•容积反馈消融是一种允许更大治疗体积的新技术。•使用容积反馈,MR 引导的超声消融治疗子宫肌瘤似乎是安全的。