Lindgren J U, Brismar B H, Wikstrom A C
Karolinska University Hospital, Department of Orthopaedics K54, Huddinge, S14186 Stockholm, Sweden.
J Bone Joint Surg Br. 2011 Oct;93(10):1427-30. doi: 10.1302/0301-620X.93B10.27645.
A 70-year-old man with an uncemented metal-on-polyethylene total hip prosthesis underwent revision arthroplasty 33 months later because of pain, swelling and recurrent dislocation. There appeared to be corrosion and metal release from the prosthetic head, resulting in pseudotumour formation and severe local soft-tissue destruction. The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven.