Suppr超能文献

使用钛钼锆铁合金柄的金属对聚乙烯全髋关节置换术中金属碎屑的不良反应。

Adverse reactions to metal debris in metal-on-polyethylene total hip arthroplasty using a titanium-molybdenum-zirconium-iron alloy stem.

作者信息

Kiran Manish, Boscainos Petros J

机构信息

Department of Trauma and Orthopaedic Surgery, NHS Tayside, Scotland, UK; Department of Trauma and Orthopaedic Surgery, School of Medicine, University of Dundee, Scotland, UK.

出版信息

J Arthroplasty. 2015 Feb;30(2):277-81. doi: 10.1016/j.arth.2014.06.030. Epub 2014 Oct 30.

Abstract

We report a series of three patients who underwent uncemented total hip arthroplasty with a modular titanium-molybdenum-zirconium-iron stem and a cobalt-chrome-molybdenum head on an ultra-high molecular weight highly cross-linked polyethylene liner bearing. All three cases subsequently developed pain and adverse reaction to metal debris, leading to revision of the implants within thirty-six months. They were subsequently found to have hypersensitivity to cobalt or chromium. However where tested, blood metal ion levels were within MHRA guideline limits. Corrosion was noted at the taper-trunnion junction. It is possible, that the multi alloy head-neck combination may lead to corrosion. Hypersensitivity to metal ions may result to ARMD at lower metal ion levels. The use of ceramic heads may help avoid this risk.

摘要

我们报告了一系列3例患者,他们接受了非骨水泥型全髋关节置换术,使用了模块化钛-钼-锆-铁柄和钴-铬-钼头,置于超高分子量高度交联聚乙烯衬垫轴承上。所有3例患者随后均出现疼痛和对金属碎屑的不良反应,导致在36个月内对植入物进行翻修。随后发现他们对钴或铬过敏。然而,在进行检测时,血液中的金属离子水平在药品和医疗产品监管局(MHRA)的指导限值范围内。在锥度-耳轴连接处发现了腐蚀。多合金头颈组合可能会导致腐蚀,这是有可能的。对金属离子过敏可能在较低金属离子水平时导致人工关节相关金属离子诱导的骨溶解(ARMD)。使用陶瓷头可能有助于避免这种风险。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验