正畸用钛合金微螺钉阳极氧化表面变化。

Surface changes of anodic oxidized orthodontic titanium miniscrew.

机构信息

Department of Orthodontics, College of Dentistry, Yonsei University, Seoul, Korea.

出版信息

Angle Orthod. 2012 May;82(3):522-8. doi: 10.2319/071311-448.1. Epub 2011 Oct 19.

Abstract

OBJECTIVE

To evaluate the structural stability of anodic oxidation treatment of miniscrews during a self-drilling procedure and an initial loading period.

MATERIALS AND METHODS

Eight orthodontic miniscrews with a machined surface and an anodic oxidized surface were placed in the mandible of two beagle dogs. With all miniscrews, an orthodontic force was applied immediately after placement and was continued for 12 weeks. After beagle dogs were sacrificed, the miniscrews were carefully removed from decalcified bone fragments. Miniscrews were evaluated by comparing and quantitatively analyzing changes in surface roughness of unused and used miniscrews (machined surface vs anodic oxidized surface) utilizing both scanning electron microscopy (SEM) and atomic force microscopy (AFM).

RESULTS

SEM revealed that only a thread edge close to the tip of the used anodic oxidized miniscrew became smooth by smearing, compared with the unused anodic oxidized miniscrew. No definite changes were observed in the thread valleys of the two groups after placement. AFM measurements demonstrated that all surface roughness parameters of thread edges of the used anodic oxidized miniscrews were significantly reduced compared with the unused anodic oxidized miniscrew (P < .05). A middle thread edge of the used anodic miniscrew surface was rougher than the unused and used machined surface miniscrews (P < .05).

CONCLUSION

Anodic oxidized miniscrews had improved surface characteristics compared with machined surface miniscrews, even if the surface texture was changed by the self-drilling procedure and during the initial loading period.

摘要

目的

评估自攻钻程序和初始加载期间阳极氧化处理的微螺钉的结构稳定性。

材料和方法

将 8 颗具有机械加工表面和阳极氧化表面的正畸微螺钉植入两只比格犬的下颌骨中。所有微螺钉在植入后立即施加正畸力,并持续 12 周。比格犬被处死,小心地从脱钙骨碎片中取出微螺钉。通过比较和定量分析未使用和使用的微螺钉(机械加工表面与阳极氧化表面)的表面粗糙度变化,利用扫描电子显微镜(SEM)和原子力显微镜(AFM)来评估微螺钉。

结果

SEM 显示,与未使用的阳极氧化微螺钉相比,只有靠近尖端的螺纹边缘因涂抹而变得光滑,而使用过的阳极氧化微螺钉的其他螺纹边缘没有明显变化。两组微螺钉在植入后,螺纹谷均未观察到明确变化。AFM 测量结果表明,与未使用的阳极氧化微螺钉相比,所有使用过的阳极氧化微螺钉的螺纹边缘表面粗糙度参数均显著降低(P <.05)。使用过的阳极氧化微螺钉的中间螺纹边缘比未使用和使用过的机械加工表面微螺钉更粗糙(P <.05)。

结论

即使在自攻钻程序和初始加载期间表面纹理发生变化,阳极氧化处理的微螺钉仍具有比机械加工表面微螺钉更好的表面特性。

相似文献

1
Surface changes of anodic oxidized orthodontic titanium miniscrew.正畸用钛合金微螺钉阳极氧化表面变化。
Angle Orthod. 2012 May;82(3):522-8. doi: 10.2319/071311-448.1. Epub 2011 Oct 19.

引用本文的文献

本文引用的文献

6
Characterization of retrieved orthodontic miniscrew implants.回收的正畸微螺钉种植体的特性分析
Am J Orthod Dentofacial Orthop. 2009 Jan;135(1):10.e1-7; discussion 10-1. doi: 10.1016/j.ajodo.2008.06.019.
9
Recommended placement torque when tightening an orthodontic mini-implant.拧紧正畸微型种植体时的推荐放置扭矩。
Clin Oral Implants Res. 2006 Feb;17(1):109-14. doi: 10.1111/j.1600-0501.2005.01211.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验