Department of Ophthalmology and Visual Science, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, Japan.
Curr Eye Res. 2012 Feb;37(2):109-14. doi: 10.3109/02713683.2011.623811. Epub 2011 Oct 26.
To investigate the influence of positional deviations on rebound tonometer measurement of intraocular pressure in humans.
Intraocular pressure was measured using the Icare rebound tonometer on the right eyes of 53 subjects in various conditions as follows: first, at a distance of 4, 6, or 8 mm from the center of the cornea with the probe perpendicular to the corneal plane; then, at 2 mm from the limbus in the nasal and temporal regions with the probe perpendicular to the corneal plane or along the visual axis; and lastly, with the angled probe touching the central cornea at angles of 10° or 20°.
Bland-Altman plots between the Goldmann applanation tonometer and rebound tonometer at various conditions revealed 95% limits of agreement ranging from ±4.5 to ±5.6 mm Hg. Intraocular pressures measured using the rebound tonometer were significantly lower than those measured using the Goldmann applanation tonometer when the rebound tonometer probe was placed 2 mm from the limbus in the temporal or nasal regions with the probe along the visual axis or when the probe was angled to touch the central cornea at an angle of 10° or 20°. In other positions, the intraocular pressures measured using the rebound tonometer were not significantly different. The rebound tonometer, noncontact tonometer, and Tonopen XL showed good agreement with the Goldmann applanation tonometer for intraocular pressure readings under optimal conditions.
The intraocular pressures determined using the rebound tonometer were approximately equal to those obtained using the Goldmann applanation tonometer when the rebound tonometer measurements were made with the probe perpendicular to the corneal plane, irrespective of the location, that is, at the central cornea or 2 mm from the limbus.
研究在人体中,位置偏差对回弹眼压计测量眼压的影响。
对 53 名受试者的右眼进行了各种条件下的眼压测量,使用 Icare 回弹眼压计,具体如下:首先,探头垂直于角膜平面,分别在距角膜中心 4、6 或 8mm 处;然后,在鼻侧和颞侧距角膜缘 2mm 处,探头垂直于角膜平面或沿视轴;最后,将倾斜探头以 10°或 20°角接触中央角膜。
在各种条件下,Goldmann 压平眼压计和回弹眼压计之间的 Bland-Altman 图显示,95%的一致性界限范围为±4.5 至±5.6mmHg。当回弹眼压计探头置于颞侧或鼻侧距角膜缘 2mm 处,探头沿视轴或探头倾斜至以 10°或 20°角接触中央角膜时,使用回弹眼压计测量的眼压明显低于使用 Goldmann 压平眼压计测量的眼压。在其他位置,使用回弹眼压计测量的眼压没有显著差异。回弹眼压计、非接触眼压计和 Tonopen XL 在最佳条件下与 Goldmann 压平眼压计对眼压读数具有良好的一致性。
当回弹眼压计测量时探头垂直于角膜平面,无论位置如何,即中央角膜或距角膜缘 2mm,使用回弹眼压计测量的眼压与使用 Goldmann 压平眼压计测量的眼压大致相等。