Barkana Yaniv, Gutfreund Shay
Department of Ophthalmology, Assaf Harofeh Medical Center, Zerifin, Israel.
Clin Exp Ophthalmol. 2014 Sep-Oct;42(7):608-14. doi: 10.1111/ceo.12272. Epub 2014 Jan 13.
Change in intraocular pressure is known to occur with body posture. Previously, markedly different estimations of this change were reported. The Icare Pro (Icare; Tiolat Oy, Helsinki, Finland) is designed to measure intraocular pressure with the subject supine. We compared the measurement of postural change in intraocular pressure obtained with this and 3 other tonometers: Goldmann applanation tonometer (CSO, Firenze, Italy), Pneumatonometer (Model 30, Reichert, Depew, NY, USA) and the Tonopen XL (Reichert).
Prospective comparative study.
Twenty-one healthy subjects.
Intraocular pressure was measured in random order with four tonometers, first sitting then after 10 min of lying.
Average postural intraocular pressure change (mmHg) and the 95% limits of agreement for measuring postural intraocular pressure change between tonometers.
Average postural intraocular pressure change (lying minus sitting, mmHg) was measured highest with the Goldmann applanation tonometer (4.1 ± 2.6) and Pneumatonometer (3.9 ± 2.1), significantly lower with the Tonopen (0.9 ± 1.7) and was negative with the Icare (-0.9 ± 3.3). The 95% limits of agreement for measuring postural intraocular pressure change between Goldmann applanation tonometer and Icare, Pneumatonometer and Tonopen were -3.2-13.3, -6.4-6.9 and -2.8-9.4 mmHg, respectively.
Postural change in intraocular pressure was measured similarly, on average, with Goldmann applanation tonometer and Pneumatonometer, and much lower with Tonopen and Icare Pro. There was poor interdevice agreement in measuring this parameter. We observed high intersubject variability, suggesting the clinical importance of this measurement in the diagnosis and management of glaucoma.
已知眼压会随身体姿势发生变化。此前,关于这种变化的估计差异显著。Icare Pro眼压计(芬兰赫尔辛基的Tiolat Oy公司生产)设计用于在受试者仰卧位时测量眼压。我们比较了使用该眼压计与其他三种眼压计测量姿势性眼压变化的情况:Goldmann压平眼压计(意大利佛罗伦萨的CSO公司生产)、气动眼压计(美国纽约州迪皮尤的Reichert公司生产的30型)和Tonopen XL眼压计(Reichert公司生产)。
前瞻性对比研究。
21名健康受试者。
使用四种眼压计以随机顺序测量眼压,先在坐位测量,然后在卧位10分钟后测量。
姿势性眼压变化的平均值(mmHg)以及各眼压计测量姿势性眼压变化的95%一致性界限。
Goldmann压平眼压计(4.1±2.6)和气动眼压计(3.9±2.1)测量的姿势性眼压变化平均值(卧位减去坐位,mmHg)最高,Tonopen眼压计(0.9±1.7)显著较低,Icare眼压计(-0.9±3.3)为负值。Goldmann压平眼压计与Icare眼压计、气动眼压计与Tonopen眼压计测量姿势性眼压变化的95%一致性界限分别为-3.2至13.3、-6.4至6.9和-2.8至9.4 mmHg。
Goldmann压平眼压计和气动眼压计测量的姿势性眼压变化平均值相近,而Tonopen眼压计和Icare Pro眼压计测量值低得多。在测量该参数时,不同仪器间的一致性较差。我们观察到受试者间变异性较高,提示该测量在青光眼诊断和管理中的临床重要性。