Technical note: a comparison of central and peripheral intraocular pressure using rebound tonometry.

作者信息

Queirós A, González-Méijome J M, Fernandes P, Jorge J, Montés-Micó R, Almeida J B, Parafita M A

机构信息

Department of Physics (Optometry), School of Sciences, University of Minho, Braga, Portugal.

出版信息

Ophthalmic Physiol Opt. 2007 Sep;27(5):506-11. doi: 10.1111/j.1475-1313.2007.00508.x.

Abstract

PURPOSE

To compare central and peripheral intraocular pressure (IOP) readings obtained with rebound tonometry.

METHODS

Intraocular pressure was measured on the right eye of 153 patients (65 males, 88 females), aged from 21 to 85 years (mean +/- S.D., 55.5 +/- 15.2 years) with the ICare rebound tonometer at centre, and 2 mm from the limbus (in the nasal and temporal regions along the 0-180 degrees corneal meridian).

RESULTS

Intraocular pressure values obtained with the ICare were 14.9 +/- 2.8; 14.1 +/- 2.5 and 14.5 +/- 2.7 mmHg at centre, nasal and temporal corneal locations, respectively. On average, nasal and temporal IOP readings were 0.75 and 0.37 mmHg lower than the central reading (p < 0.05 and p > 0.05, respectively). A highly significant correlation was found between central and peripheral measurements in nasal (r(2) = 0.905; p < 0.001) and temporal (r(2) = 0.879; p < 0.001) regions along the horizontal meridian. Almost 80% of patients presented nasal IOP values within +/-1 mmHg of the central value.

CONCLUSIONS

Intraocular pressure values measured with the ICare rebound tonometer on the nasal corneal region is slightly lower on average and highly correlated with IOP values recorded at corneal centre. Both nasal and temporal readings are in good agreement with central IOP, and could be used to obtain a reliable estimate of rebound IOP in corneas where central readings cannot be taken.

摘要

目的

比较使用回弹式眼压计获得的中央和周边眼压读数。

方法

使用ICare回弹式眼压计对153例患者(65例男性,88例女性)的右眼进行眼压测量,年龄在21至85岁之间(平均±标准差,55.5±15.2岁),测量位置分别为角膜中央、距角膜缘2mm处(沿角膜0 - 180度子午线的鼻侧和颞侧区域)。

结果

使用ICare眼压计在角膜中央、鼻侧和颞侧测量得到的眼压值分别为14.9±2.8、14.1±2.5和14.5±2.7 mmHg。平均而言,鼻侧和颞侧的眼压读数分别比中央读数低0.75和0.37 mmHg(p分别<0.05和p>0.05)。在水平子午线上的鼻侧(r² = 0.905;p < 0.001)和颞侧(r² = 0.879;p < 0.001)区域,中央和周边测量值之间存在高度显著的相关性。几乎80%的患者鼻侧眼压值在中央值的±1 mmHg范围内。

结论

使用ICare回弹式眼压计在鼻侧角膜区域测量的眼压值平均略低,且与角膜中央记录的眼压值高度相关。鼻侧和颞侧读数与中央眼压值都高度一致,可用于在无法获取中央读数的角膜上可靠地估计回弹眼压。