Franz M R, Chin M C, Sharkey H R, Griffin J C, Scheinman M M
Cardiology Division, Stanford University School of Medicine, California 94305.
J Am Coll Cardiol. 1990 Oct;16(4):878-86. doi: 10.1016/s0735-1097(10)80336-5.
In vivo correlations of action potential duration measured by a monophasic action potential catheter and effective refractory period measured by a separate pacing catheter have been poor, probably because of the known variability of both action potential duration and effective refractory period between different ventricular sites. In this study, a new quadripolar contact electrode catheter designed for simultaneous pacing and monophasic action potential recording at closely adjacent sites (2 mm separation between recording electrodes and pacing electrodes) was tested in five closed chest dogs and four patients. Dog studies: Pacing thresholds were extremely low, ranging from 0.02 to 0.25 mA (mean +/- SD 0.099 +/- 0.051, n = 36) and were stable over time (less than 20% increase during 1 h of continuous pacing). Because of the close proximity of pacing and recording electrodes, the pacing artifact nearly coincided with the monophasic action potential upstroke. Because of the low pacing threshold, however, pacing artifacts were small (33 +/- 17% of the monophasic action potential amplitude at twice diastolic threshold strength) and did not affect the duration or configuration of the simultaneously recorded monophasic action potential. The short stimulus response time and the undisturbed monophasic action potential signal fidelity during pacing allowed precise simultaneous measurements of action potential duration and effective refractory period at the same endocardial site.(ABSTRACT TRUNCATED AT 250 WORDS)
通过单相动作电位导管测量的动作电位持续时间与通过单独的起搏导管测量的有效不应期之间的体内相关性一直较差,这可能是因为已知不同心室部位之间动作电位持续时间和有效不应期存在变异性。在本研究中,一种新的四极接触电极导管被设计用于在紧密相邻的部位同时进行起搏和单相动作电位记录(记录电极和起搏电极之间相距2 mm),并在5只开胸犬和4例患者中进行了测试。犬类研究:起搏阈值极低,范围为0.02至0.25 mA(平均值±标准差0.099±0.051,n = 36),并且随时间稳定(在连续起搏1小时内增加不到20%)。由于起搏电极和记录电极距离很近,起搏伪迹几乎与单相动作电位的上升支重合。然而,由于起搏阈值低,起搏伪迹较小(在两倍舒张阈值强度时为单相动作电位幅度的33±17%),并且不影响同时记录的单相动作电位的持续时间或形态。起搏期间短的刺激反应时间和未受干扰的单相动作电位信号保真度允许在同一心内膜部位精确同时测量动作电位持续时间和有效不应期。(摘要截断于250字)