Shao Xiaodong, Li Ying, Li Fagen, Liu Yangqin, Song Zhenghua
CNPC Tubular Goods Research Institute, No. 32 Dianzier Rd, Xi'an 710065, People's Republic of China.
J AOAC Int. 2011 Sep-Oct;94(5):1461-6. doi: 10.5740/jaoacint.10-175.
A novel flow injection chemiluminescence method with a myoglobin-luminol system is described for determining aniracetam. Myoglobin-bound aniracetam produced a complex that catalyzed the chemiluminescence reaction between luminol and myoglobin, leading to fast chemiluminescence. The chemiluminescence intensity in the presence of aniracetam was remarkably enhanced compared with that in the absence of aniracetam. Under the optimum reaction conditions the chemiluminescence increment produced was proportional to the concentration of aniracetam in the range of 0.1-1000.0 ng/mL (R2 = 0.9992), with a detection limit of 0.03 ng/mL (3delta). At a flow rate of 2.0 mL/min, the whole process, including sampling and washing, could be completed in 0.5 min, offering a sampling efficiency of 120/h; the RSD was less than 3.0% (n = 5). The method was satisfactory for determination of aniracetam in pharmaceutical preparations and human urine and serum samples. A possible mechanism of the reaction is also discussed.