BACKGROUND

A new simple technique based on iontophoresis technology (EZSCAN, Impeto Medical, Paris, France) has recently been developed for the screening of diabetes. In the present study, we investigated the accuracy of this system for the diagnosis of diabetes mellitus in Chinese.

METHODS

We performed the EZSCAN test in diabetic and non-diabetic subjects. EZSCAN measures electrochemical conductance (EC) at forehead, hands and feet, and derives a diabetes index with a value ranging from 0 to 100. Diabetes mellitus was defined as a plasma glucose concentration of at least 7 mmol/l at fasting or 11.1 mmol/l at 2 hours after glucose load, or as the use of antidiabetic drugs.

RESULTS

The 195 study participants (51% men, mean age 52 years) included 75 diabetic patients (use of antidiabetic drugs 81%) and 120 non-diabetic subjects. EC (micro Siemens, μSi) was significantly (P < 0.001) lower in diabetic patients at the hands (44 vs. 61) and feet (51 vs. 69) locations, but not at the forehead (15 vs. 17, P = 0.39). When a diabetes index of 40 (suggested by the manufacturer) was used as the threshold, the sensitivity and specificity for the diagnosis of diabetes mellitus was 85% and 64%, respectively. In 80 patients who underwent an oral glucose tolerance test, EC at hands and feet and the diabetes index were significantly (P < 0.001) associated with both 2-hour post-load plasma glucose and serum glycosylated haemoglobin.

CONCLUSIONS

EZSCAN might be useful in screening diabetes mellitus with reasonable sensitivity and specificity.

摘要

背景

最近,一种基于电渗透技术的新的简单技术(EZSCAN,Impeto Medical,巴黎,法国)已被开发用于糖尿病的筛查。在本研究中,我们研究了该系统在中国人群中诊断糖尿病的准确性。

方法

我们对糖尿病患者和非糖尿病患者进行了 EZSCAN 测试。EZSCAN 测量额、手和脚的电化学传导(EC),并得出一个范围在 0 到 100 之间的糖尿病指数。糖尿病定义为空腹时血糖浓度至少为 7mmol/L 或葡萄糖负荷后 2 小时血糖浓度至少为 11.1mmol/L,或使用降糖药物。

结果

195 名研究参与者(51%为男性,平均年龄 52 岁)包括 75 名糖尿病患者(使用降糖药物 81%)和 120 名非糖尿病患者。糖尿病患者的手部(44 比 61)和足部(51 比 69)EC(微西门子,μSi)显著降低(P < 0.001),但额部无显著差异(15 比 17,P = 0.39)。当制造商建议的 40 糖尿病指数作为阈值时,该方法诊断糖尿病的敏感性和特异性分别为 85%和 64%。在 80 名接受口服葡萄糖耐量试验的患者中,手部和足部的 EC 以及糖尿病指数与 2 小时餐后血糖和血清糖化血红蛋白均显著相关(P < 0.001)。

结论

EZSCAN 可能具有合理的敏感性和特异性,可用于筛查糖尿病。