Welk Aaron R, Baumgartner J Craig, Marshall J Gordon
Department of Endodontology, Oregon Health & Sciences University, School of Dentistry, Portland 97201, USA.
J Endod. 2003 Aug;29(8):497-500. doi: 10.1097/00004770-200308000-00002.
The purpose of this study was to compare the accuracy of a two-frequency (Root ZX) and a five-frequency (Endo Analyzer Model 8005) electronic apex locator under clinical conditions. Thirty-two teeth planned for extraction were used. The coronal portion of each canal was flared using Gates Glidden drills and Orifice Shapers. The canals were irrigated with 2.6% sodium hypochlorite. A K-type file was used to determine a separate working length in each canal using the electronic apex locators. The teeth were extracted and the apical 4 mm of each root canal was exposed along the long axis of the tooth. Photographic slides of each canal were projected and the file position in relation to the minor diameter was determined by two investigators. The mean distance between the electronic apex locator working length and minor diameter was 1.03 mm for the Endo Analyzer and 0.19 mm for the Root ZX. A paired sample t test showed that the Endo Analyzer had significantly longer readings beyond the minor diameter than the Root ZX (p < 0.0001). The ability to locate the minor diameter (+/- 0.5 mm) was 90.7% for the Root ZX and 34.4% for the Endo Analyzer Model 8005.
本研究的目的是比较临床条件下双频(Root ZX)和五频(Endo Analyzer Model 8005)电子根尖定位仪的准确性。使用了32颗计划拔除的牙齿。每个根管的冠部使用Gates Glidden钻和根管口成形器进行扩管。根管内用2.6%的次氯酸钠冲洗。使用K型锉,通过电子根尖定位仪确定每个根管的独立工作长度。拔除牙齿后,沿牙齿长轴暴露每个根管根尖4mm的部分。对每个根管的幻灯片进行投影,由两名研究人员确定锉相对于根管最小直径的位置。Endo Analyzer电子根尖定位仪工作长度与根管最小直径之间的平均距离为1.03mm,Root ZX为0.19mm。配对样本t检验显示,Endo Analyzer在根管最小直径之外的读数显著长于Root ZX(p < 0.0001)。Root ZX定位根管最小直径(±0.5mm)的能力为90.7%,Endo Analyzer Model 8005为34.4%。