Radiation Physics Department, Queen's Centre for Oncology and Haematology, Hull, UK.
Br J Radiol. 2012 Sep;85(1017):e630-9. doi: 10.1259/bjr/47377285. Epub 2012 Jan 17.
The purpose of this study was to derive an optimum radiographic technique for computed radiography (CR) chest imaging using a digitally reconstructed radiograph computer simulator. The simulator is capable of producing CR chest radiographs of adults with various tube potentials, receptor doses and scatter rejection.
Four experienced image evaluators graded images of average and obese adult patients at different potentials (average-sized, n=50; obese, n=20), receptor doses (n=10) and scatter rejection techniques (average-sized, n=20; obese, n=20). The quality of the images was evaluated using visually graded analysis. The influence of rib contrast was also assessed.
For average-sized patients, image quality improved when tube potential was reduced compared with the reference (102 kVp). No scatter rejection was indicated. For obese patients, it has been shown that an antiscatter grid is indicated, and should be used in conjunction with as low a tube potential as possible (while allowing exposure times <20 ms). It is also possible to reduce receptor air kerma by 50% without adversely influencing image quality. Rib contrast did not interfere at any tube potential.
A virtual clinical trial has been performed with simulated chest CR images. Results indicate that low tube potentials (<102 kVp) are optimal for average and obese adults, the former acquired without scatter rejection, the latter with an anti-scatter grid. Lower receptor (and therefore patient doses) than those used clinically are possible while maintaining adequate image quality.
本研究旨在利用数字重建射线照片计算机模拟器为计算机射线照相(CR)胸部成像制定最佳射线照相技术。该模拟器能够生成具有不同管电压、接收器剂量和散射抑制的成人 CR 胸部射线照片。
四位经验丰富的图像评估员对不同管电压(普通体型,n=50;肥胖体型,n=20)、接收器剂量(n=10)和散射抑制技术(普通体型,n=20;肥胖体型,n=20)下的普通体型和肥胖成年患者的图像进行了分级。使用视觉分级分析评估图像质量。还评估了肋骨对比度的影响。
对于普通体型患者,与参考(102 kVp)相比,降低管电压可提高图像质量。不建议使用散射抑制。对于肥胖患者,表明需要使用散射抑制栅格,并且应尽可能结合使用尽可能低的管电压(同时允许曝光时间<20ms)。也可以将接收器空气比释动能降低 50%,而不会对图像质量产生不利影响。在任何管电压下,肋骨对比度都不会产生干扰。
已经对模拟胸部 CR 图像进行了虚拟临床试验。结果表明,对于普通和肥胖成年人,低管电压(<102 kVp)是最佳选择,前者无需散射抑制即可获得,后者则需要使用散射抑制栅格。在保持足够的图像质量的同时,可以使用比临床应用低的接收器(因此也低患者剂量)。