Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria.
Graefes Arch Clin Exp Ophthalmol. 2012 Sep;250(9):1261-6. doi: 10.1007/s00417-012-1946-3. Epub 2012 Feb 25.
Several methods have been proposed for measuring macular pigment optical density (MPOD). To date, none of the realized techniques can be considered as "gold standard". A key issue for the clinical applicability of a method is its repeatability. In this study, we investigated short- and mid-term repeatability of MPOD measurements using reflectometry.
A total of 12 healthy young subjects were measured 5 times on 5 consecutive days. Repeatability over 6 months was investigated in patients with AMD. The data in AMD patients were taken from a recently published placebo controlled study investigating the effect of lutein supplementation on MPOD (n = 37; [1]). Four measurements over 6 months were used to calculate repeatability. Spectral fundus reflectance of the fovea was measured in a 2.3° detection field with a custom-built fundus reflectometer. Calculation of MPOD was based on a previously published fundus reflectance model.
The coefficients of variation were 6.2 ± 2.4% and 8.0 ± 5.5% in the healthy and AMD group, respectively. Bland-Altman plots indicate that the difference between measurements at day 1 and day 5 in healthy subjects and day 1 and month 6 in AMD patients was small. The maximum deviation in MPOD in a healthy subject was 0.07 (22.6%), and 0.17 (51.5%) in a patient with AMD.
Reflectometry provides adequate short-term and mid-term repeatability for measuring MPOD. Accordingly, the technique makes it possible to monitor MPOD in patients with AMD and to study the influence of supplementation in these subjects. In addition, the variability of the technique is small enough to allow for clinical trials with reasonable sample size.
已有多种方法被提出用于测量黄斑色素光学密度(MPOD)。迄今为止,尚无一种已实现的技术可被视为“金标准”。对于方法的临床适用性,一个关键问题是其可重复性。在这项研究中,我们使用反射计研究了 MPOD 测量的短期和中期重复性。
总共 12 名健康的年轻受试者在连续 5 天内进行了 5 次测量。在 AMD 患者中研究了 6 个月的重复性。AMD 患者的数据取自最近发表的一项安慰剂对照研究,该研究调查了叶黄素补充对 MPOD 的影响(n = 37;[1])。使用 6 个月内的 4 次测量来计算重复性。使用定制的眼底反射计在 2.3°检测视野中测量黄斑区的光谱眼底反射。MPOD 的计算基于先前发表的眼底反射模型。
健康组和 AMD 组的变异系数分别为 6.2 ± 2.4%和 8.0 ± 5.5%。Bland-Altman 图表明,健康受试者第 1 天和第 5 天之间以及 AMD 患者第 1 天和第 6 个月之间测量值的差异较小。健康受试者的 MPOD 最大偏差为 0.07(22.6%),AMD 患者为 0.17(51.5%)。
反射计为测量 MPOD 提供了足够的短期和中期重复性。因此,该技术使得在 AMD 患者中监测 MPOD 并研究这些患者补充的影响成为可能。此外,该技术的变异性足够小,可以进行具有合理样本量的临床试验。