腕戴加速度计在儿童中的有效性和可比性。

Validity and comparability of a wrist-worn accelerometer in children.

机构信息

Åstrand Laboratory of Work Physiology, The Swedish School of Sport and Health Sciences, Stockholm, Sweden.

出版信息

J Phys Act Health. 2012 Mar;9(3):389-93.

Abstract

BACKGROUND

Wrist-worn accelerometers may provide an alternative to hip-worn monitors for assessing physical activity as they are easier to wear and may thus facilitate long-term recordings. The current study aimed at a) assessing the validity of the Actiwatch (wrist-worn) for estimating energy expenditure, b) determining cut-off values for light, moderate, and vigorous activities, c) studying the comparability between the Actiwatch and the Actigraph (hip-worn), and d) assessing reliability.

METHODS

For validity, indirect calorimetry was used as criterion measure. ROC-analyses were applied to identify cut-off values. Comparability was tested by simultaneously wearing of the 2 accelerometers during free-living condition. Reliability was tested in a mechanical shaker.

RESULTS

All-over correlation between accelerometer output and energy expenditure were found to be 0.80 (P < .001).Based on ROC-analysis, cut-off values for 1.5, 3, and 6 METs were found to be 80, 262, and 406 counts per 15 s, respectively. Energy expenditure estimates differed between the Actiwatch and the Actigraph (P < .05). The intra- and interinstrument coefficient of variation of the Actiwatch ranged between 0.72% and 8.4%.

CONCLUSION

The wrist-worn Actiwatch appears to be valid and reliable for estimating energy expenditure and physical activity intensity in children aged 8 to 10 years.

摘要

背景

腕戴式加速度计在评估身体活动方面可能替代髋戴式监测器,因为它们更容易佩戴,因此可以促进长期记录。本研究旨在:a)评估 Actiwatch(腕戴式)估算能量消耗的有效性,b)确定轻、中、高强度活动的截断值,c)研究 Actiwatch 和 Actigraph(髋戴式)之间的可比性,d)评估可靠性。

方法

间接热量测定法作为标准测量法用于评估有效性。应用 ROC 分析确定截断值。通过在自由生活条件下同时佩戴 2 个加速度计来测试可比性。在机械振动台上测试可靠性。

结果

发现加速度计输出与能量消耗之间的总体相关性为 0.80(P <.001)。基于 ROC 分析,1.5、3 和 6 MET 的截断值分别为 80、262 和 406 计数/15 秒。Actiwatch 和 Actigraph 之间的能量消耗估计值存在差异(P <.05)。Actiwatch 的仪器内和仪器间变异系数在 0.72%到 8.4%之间。

结论

腕戴式 Actiwatch 似乎在评估 8 至 10 岁儿童的能量消耗和身体活动强度方面是有效和可靠的。