Evaluation of the PotoClean(®) decontamination technology for reprocessing of water supply lines in dental units during routine work.

作者信息

Kramer Axel, Koburger Torsten, Taube Lisa-Dorothea, Menzel Michael, Meyer Georg, Assadian Ojan

机构信息

Institute of Hygiene and Environmental Medicine, University Medicine Greifswald, Germany.

出版信息

GMS Krankenhhyg Interdiszip. 2012;7(1):Doc10. doi: 10.3205/dgkh000194. Epub 2012 Apr 4.

Abstract

BACKGROUND

A frequent problem in dental units is the microbial contamination of water and biofilm formation in the water supply lines. After random identification of a bacterial contaminated dental unit (310 cfu/ml) in a practise with 3 dental units we implemented the present study to evaluate the efficacy of the PotoClean(®) technology, based on anodic oxidation.

METHOD

The efficacy of a regular low concentrated permanent decontamination (1 mg Cl/L) with an additional intensive decontamination by PotoClean(®) (three times 20 mg Cl/ml for 2 h) on three dental units was tested over 7 months. Microbial contamination, total chlorine concentration and redox potential have been analyzed. Dental unit A and B was 15 years old, unit C 5 years.

RESULTS

After 3 intensive decontaminations, in dental unit A and B the number of bacteria and moulds could be reduced less than 7 d. Thereafter the bacteria counts increased again during the subsequent 7 month period and the amount of moulds was with some exceptions 300 cfu/ml, although PotoClean(®) was constantly added in the system (1 mg Cl/L). After further 7.5 month only with low concentrated permanent disinfection (1 mg Cl/L) both units were successful decontaminated. Dental unit C represented an object which was easier to decontaminate because of the advanced construction (prevention of water stagnation) and the shorter useful life. At the beginning of the decontamination it was no bacterial contamination, but moulds were contained (300 cfu/ml). Already after the first intensive decontamination, no further bacteria and moulds could be detected.

DISCUSSION

An important factor for the efficacy of PotoClean(®) was the age of the units and their construction. For a new generation of dental units PotoClean(®) was effective during the whole period of monitoring. For two old types of dental unit with massive biofilm development the successful decontamination needed more than 7 month.

CONCLUSION

The PotoClean(®) technology has resulted in even old-type turbines with intensive biofilm formation to complete decontamination. In a recent turbine design already after the first intensive decontamination with PotoClean(®) and its continuous use (1 mg Cl/L) no more contamination by bacteria and moulds were detectable.

摘要
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/534e/3334947/c912ee237ddb/KHI-07-10-g-001.jpg

相似文献

3
Purge- and intensive-purge decontamination of dental units contaminated with biofilm.
GMS Krankenhhyg Interdiszip. 2012;7(1):Doc11. doi: 10.3205/dgkh000195. Epub 2012 Apr 4.
4
Dental unit water: bacterial decontamination of old and new dental units by flushing water.
Int J Dent Hyg. 2008 Feb;6(1):56-62. doi: 10.1111/j.1601-5037.2007.00278.x.
6
A pilot study of three methods for the reduction of bacterial contamination of dental unit water systems in routine use.
Int J Hyg Environ Health. 2002 Feb;204(5-6):303-8. doi: 10.1078/1438-4639-00120.
7
Monitoring the decontamination efficacy of the novel Poseidon-S disinfectant system in dental unit water lines.
J Microbiol Immunol Infect. 2017 Jun;50(3):270-276. doi: 10.1016/j.jmii.2015.05.006. Epub 2015 May 14.
8
Formation and decontamination of biofilms in dental unit waterlines.
J Periodontol. 2003 Nov;74(11):1595-609. doi: 10.1902/jop.2003.74.11.1595.
9
In vitro modeling of dental water line contamination and decontamination.
FEMS Microbiol Lett. 2004 Jun 15;235(2):363-7. doi: 10.1016/j.femsle.2004.05.006.
10
Effects of iodine in microbial control of dental treatment water.
J Contemp Dent Pract. 2011 May 1;12(3):143-51. doi: 10.5005/jp-journals-10024-1025.

引用本文的文献

1
Efficacy of BRS and Alpron/Bilpron Disinfectants for Dental Unit Waterlines: A Six-Year Study.
Int J Environ Res Public Health. 2020 Apr 12;17(8):2634. doi: 10.3390/ijerph17082634.

本文引用的文献

4
Analysis of transmission pathways of Pseudomonas aeruginosa between patients and tap water outlets.
Crit Care Med. 2002 Oct;30(10):2222-8. doi: 10.1097/00003246-200210000-00008.
6
Multiparametric analysis of waterline contamination in dental units.
Appl Environ Microbiol. 1996 Nov;62(11):3954-9. doi: 10.1128/aem.62.11.3954-3959.1996.
7
Legionella pneumonia in transplant recipients: a cluster of cases of eight years' duration.
J Hosp Infect. 1994 Mar;26(3):191-202. doi: 10.1016/0195-6701(94)90042-6.
8
Prevalence of Legionella-specific IgG and IgM antibody in a dental clinic population.
J Dent Res. 1985 Dec;64(12):1382-5. doi: 10.1177/00220345850640121101.
9
Sink flora in a long-stay hospital is determined by the patients' oral and rectal flora.
Epidemiol Infect. 1989 Apr;102(2):231-8. doi: 10.1017/s0950268800029903.
10
Influence of biofilms on microbial contamination in dental unit water.
J Dent. 1991 Oct;19(5):290-5. doi: 10.1016/0300-5712(91)90075-a.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验