Department of Ophthalmology, Seoul National University College of Medicine, Seoul, Korea.
Invest Ophthalmol Vis Sci. 2012 Jul 1;53(8):4347-53. doi: 10.1167/iovs.12-9673.
To investigate the diagnostic ability of posterior pole asymmetry analysis (PPAA) with spectral domain optical coherence tomography (SD-OCT) for detecting localized retinal nerve fiber layer defects (RNFLD).
Eighty-four open-angle glaucoma subjects (84 eyes) with localized, wedge-shape RNFLDs by red-free RNFL photography and 122 eyes of healthy subjects were enrolled. The subjects were examined by SD-OCT to obtain circumpapillary RNFL (cpRNFL) thickness as well as PPAA. The PPAA provides a corresponding cell to cell comparison between hemispheres within the central 20° and presents the difference using gray scale. Area under the receiver operating characteristic curve (AUROC) was calculated. The sensitivity and specificity of PPAA to detect glaucoma with the localized defects were also examined using various criteria; two consecutive black cells (criterion A), three consecutive black cells (criterion B), and four consecutive black cells (criterion C).
An AUROC of PPAA based on the number of black cells was 0.958 ± 0.013. The sensitivity and specificity of PPAA were 95.2% and 81.1% for criterion A, 83.3% and 92.6% for criterion B, and 69.0% and 98.4% for criterion C, respectively. The sensitivity and specificity of the cpRNFL thickness were 85.7% and 95.1%, respectively.
SD-OCT PPAA detects localized RNFLDs with high sensitivity and specificity, which are comparable to the cpRNFL thickness profile of OCT. These findings suggest that the PPAA can be complementary to other tests for diagnosing glaucoma in patients with localized RNFLDs.
利用频域光相干断层扫描(SD-OCT)分析后极部不对称(PPAA)来检测局限性视网膜神经纤维层缺损(RNFLD)。
纳入 84 例局限性、楔形 RNFLD 的开角型青光眼患者(84 只眼)和 122 例健康对照者,采用眼底自发荧光(RNFL)摄影测量其局限性 RNFLD,并用 SD-OCT 获得周边视网膜神经纤维层(cpRNFL)厚度和 PPAA。PPAA 提供了中央 20°范围内半球间逐点逐行的对应比较,并以灰度表示差异。计算受试者工作特征曲线下面积(AUROC)。还采用不同标准(连续两个黑色细胞标准 A、连续三个黑色细胞标准 B 和连续四个黑色细胞标准 C)来检测 PPAA 对局限性病变青光眼的诊断灵敏度和特异性。
基于黑色细胞数的 PPAA AUROC 为 0.958±0.013。PPAA 的灵敏度和特异性分别为标准 A 时的 95.2%和 81.1%、标准 B 时的 83.3%和 92.6%,以及标准 C 时的 69.0%和 98.4%。cpRNFL 厚度的灵敏度和特异性分别为 85.7%和 95.1%。
SD-OCT 的 PPAA 检测局限性 RNFLD 的灵敏度和特异性较高,与 OCT 的 cpRNFL 厚度图谱相当。这些结果表明,PPAA 可以与其他检测手段互补,用于诊断局限性 RNFLD 患者的青光眼。