Na Jung Hwa, Sung Kyung Rim, Lee Youngrok
Division of Biostatistics, Center for Medical Research and Information, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
Korean J Ophthalmol. 2012 Jun;26(3):169-73. doi: 10.3341/kjo.2012.26.3.169. Epub 2012 May 22.
The aim of this study was to investigate factors associated with the signal strengths (SS, image quality scores) of optic disc and macular images obtained using Cirrus spectral domain optical coherence tomography (OCT).
Ninety-two glaucomatous eyes were imaged using the Cirrus OCT macular and optic disc cube modes after pupil dilation. The influences of patient age, spherical equivalent, cataract presence, and cataract and glaucoma severity (visual field mean deviation), on the SS of images obtained using the two cube modes were compared between patients whose images showed high SS (SS ≥7) and low SS (SS <7).
The signal strength was significantly higher in images obtained using the macular cube compared to the optic disc cube mode (7.8 ± 1.3 vs. 6.9 ± 1.1, respectively; p = 0.001). Age and visual acuity of patients differed significantly between the high- and low-SS groups when data acquired using the optic disc (p = 0.027 and 0.012, respectively) and macular cube modes (p = 0.046 and 0.014, respectively) were analyzed. When the optic disc cube mode was employed, the extent of cataracts was significantly related to SS, whereas when the macular cube mode was used, none of the factors analyzed was significantly associated with SS.
Age, visual acuity, and the extent of cataracts were significantly associated with images of higher SS when the Cirrus OCT optic disc cube mode was employed.
本研究旨在调查与使用Cirrus光谱域光学相干断层扫描(OCT)获得的视盘和黄斑图像的信号强度(SS,图像质量评分)相关的因素。
92只青光眼患眼在散瞳后使用Cirrus OCT黄斑和视盘立方模式进行成像。比较了图像显示高信号强度(SS≥7)和低信号强度(SS<7)的患者中,患者年龄、等效球镜度、白内障的存在以及白内障和青光眼严重程度(视野平均偏差)对使用两种立方模式获得的图像信号强度的影响。
与视盘立方模式相比,使用黄斑立方模式获得的图像信号强度显著更高(分别为7.8±1.3和6.9±1.1;p=0.001)。在对视盘(分别为p=0.027和0.012)和黄斑立方模式(分别为p=0.046和0.014)采集的数据进行分析时,高信号强度组和低信号强度组患者的年龄和视力存在显著差异。当采用视盘立方模式时,白内障的程度与信号强度显著相关,而当使用黄斑立方模式时,所分析的因素均与信号强度无显著关联。
当采用Cirrus OCT视盘立方模式时,年龄、视力和白内障程度与更高信号强度的图像显著相关。