Effect of signal strength on reproducibility of circumpapillary retinal nerve fiber layer thickness measurement and its classification by spectral-domain optical coherence tomography.

机构信息

Department of Ophthalmology, Institute of Vision Research, Yonsei University College of Medicine, CPO Box 8044, Seoul, 120-752, Korea.

Department of Applied Statistics, Yonsei University, Seoul, Korea.

出版信息

Jpn J Ophthalmol. 2011 May;55(3):220-227. doi: 10.1007/s10384-011-0023-5. Epub 2011 May 11.

Abstract

PURPOSE

To assess the effect of signal strength (SS) on the reproducibility of circumpapillary retinal nerve fiber layer (cpRNFL) thickness measurement (measurement agreement) and its color-coded classification (classification agreement) by Cirrus spectral-domain optical coherence tomography (OCT).

METHODS

Seven hundred and eighty-nine participants, consecutively enrolled from December 2009 to August 2010, underwent two repeated Cirrus OCT scans to measure cpRNFL thickness (optic disc cube 200 × 200). Intraclass correlation (ICC) and linear-weighted kappa coefficient (κ) were calculated as indicators for cpRNFL measurement and classification agreement. The difference in cpRNFL thickness measurements between repeated OCT scans (inter-scan measurement difference) was correlated with mean SS, SS change or difference between repeated OCT scans (inter-scan SS difference), and average cpRNFL thickness.

RESULTS

Mean SS was negatively correlated with inter-scan SS difference and was positively correlated with the inter-scan measurement difference. Repeated scans with SS change (inter-scan SS difference = 1 or 2) showed larger variability than scans without SS change (inter-scan SS difference = 0) in total average cpRNFL and all quadrant maps except for the nasal quadrant. Multivariate analysis revealed that inter-scan measurement differences increased in subjects with thinner cpRNFL thicknesses and in scans with lower mean SS and higher inter-scan SS differences. Measurement agreement (ICC) in groups with larger inter-scan SS differences (=2) was lower than those with smaller inter-scan SS differences (=0 or 1) in their average cpRNFL and inferior quadrant maps. Classification agreement (κ) in groups with larger inter-scan SS differences (=2) was lower than those with smaller inter-scan SS differences (=0 or 1) in their average cpRNFL and quadrant maps (superior and inferior quadrant) although statistically not significant.

CONCLUSIONS

Special attention should be paid when comparing the cpRNFL thickness measurement and color-coded classification of 2 or more Cirrus OCT scans as the agreements may be susceptible to SS differences.

摘要

目的

评估信号强度(SS)对 Cirrus 频域光学相干断层扫描(OCT)检测的周边视网膜神经纤维层(cpRNFL)厚度测量(测量一致性)及其彩色编码分类(分类一致性)的重复性的影响。

方法

2009 年 12 月至 2010 年 8 月连续纳入 789 名参与者,对其进行两次 Cirrus OCT 扫描以测量 cpRNFL 厚度(视盘立方 200×200)。采用组内相关系数(ICC)和线性加权 Kappa 系数(κ)评估 cpRNFL 测量和分类一致性。两次 OCT 扫描间 cpRNFL 厚度测量差值(扫描间测量差值)与平均 SS、SS 变化或两次 OCT 扫描间差值(扫描间 SS 差值)和平均 cpRNFL 厚度相关。

结果

平均 SS 与扫描间 SS 差值呈负相关,与扫描间测量差值呈正相关。SS 变化(扫描间 SS 差值=1 或 2)的重复扫描比 SS 无变化(扫描间 SS 差值=0)的总平均 cpRNFL 和除鼻侧象限外所有象限图的变异性更大。多变量分析显示,在 cpRNFL 较薄和平均 SS 较低、扫描间 SS 差值较高的患者中,扫描间测量差值增加。在 SS 差值较大(=2)的组中,测量一致性(ICC)在平均 cpRNFL 和下象限图中低于 SS 差值较小(=0 或 1)的组;在 SS 差值较大(=2)的组中,分类一致性(κ)在平均 cpRNFL 和象限图(上、下象限)中低于 SS 差值较小(=0 或 1)的组,尽管统计学上无显著差异。

结论

当比较 2 次或更多次 Cirrus OCT 扫描的 cpRNFL 厚度测量和彩色编码分类时,应特别注意 SS 差异可能会影响其一致性。