To assess the effect of signal strength (SS) on the reproducibility of circumpapillary retinal nerve fiber layer (cpRNFL) thickness measurement (measurement agreement) and its color-coded classification (classification agreement) by Cirrus spectral-domain optical coherence tomography (OCT).
Seven hundred and eighty-nine participants, consecutively enrolled from December 2009 to August 2010, underwent two repeated Cirrus OCT scans to measure cpRNFL thickness (optic disc cube 200 × 200). Intraclass correlation (ICC) and linear-weighted kappa coefficient (κ) were calculated as indicators for cpRNFL measurement and classification agreement. The difference in cpRNFL thickness measurements between repeated OCT scans (inter-scan measurement difference) was correlated with mean SS, SS change or difference between repeated OCT scans (inter-scan SS difference), and average cpRNFL thickness.
Mean SS was negatively correlated with inter-scan SS difference and was positively correlated with the inter-scan measurement difference. Repeated scans with SS change (inter-scan SS difference = 1 or 2) showed larger variability than scans without SS change (inter-scan SS difference = 0) in total average cpRNFL and all quadrant maps except for the nasal quadrant. Multivariate analysis revealed that inter-scan measurement differences increased in subjects with thinner cpRNFL thicknesses and in scans with lower mean SS and higher inter-scan SS differences. Measurement agreement (ICC) in groups with larger inter-scan SS differences (=2) was lower than those with smaller inter-scan SS differences (=0 or 1) in their average cpRNFL and inferior quadrant maps. Classification agreement (κ) in groups with larger inter-scan SS differences (=2) was lower than those with smaller inter-scan SS differences (=0 or 1) in their average cpRNFL and quadrant maps (superior and inferior quadrant) although statistically not significant.
Special attention should be paid when comparing the cpRNFL thickness measurement and color-coded classification of 2 or more Cirrus OCT scans as the agreements may be susceptible to SS differences.
评估信号强度(SS)对 Cirrus 频域光学相干断层扫描(OCT)检测的周边视网膜神经纤维层(cpRNFL)厚度测量(测量一致性)及其彩色编码分类(分类一致性)的重复性的影响。
2009 年 12 月至 2010 年 8 月连续纳入 789 名参与者,对其进行两次 Cirrus OCT 扫描以测量 cpRNFL 厚度(视盘立方 200×200)。采用组内相关系数(ICC)和线性加权 Kappa 系数(κ)评估 cpRNFL 测量和分类一致性。两次 OCT 扫描间 cpRNFL 厚度测量差值(扫描间测量差值)与平均 SS、SS 变化或两次 OCT 扫描间差值(扫描间 SS 差值)和平均 cpRNFL 厚度相关。
平均 SS 与扫描间 SS 差值呈负相关,与扫描间测量差值呈正相关。SS 变化(扫描间 SS 差值=1 或 2)的重复扫描比 SS 无变化(扫描间 SS 差值=0)的总平均 cpRNFL 和除鼻侧象限外所有象限图的变异性更大。多变量分析显示,在 cpRNFL 较薄和平均 SS 较低、扫描间 SS 差值较高的患者中,扫描间测量差值增加。在 SS 差值较大(=2)的组中,测量一致性(ICC)在平均 cpRNFL 和下象限图中低于 SS 差值较小(=0 或 1)的组;在 SS 差值较大(=2)的组中,分类一致性(κ)在平均 cpRNFL 和象限图(上、下象限)中低于 SS 差值较小(=0 或 1)的组,尽管统计学上无显著差异。
当比较 2 次或更多次 Cirrus OCT 扫描的 cpRNFL 厚度测量和彩色编码分类时,应特别注意 SS 差异可能会影响其一致性。