Rey Markus, Weber Edgar W, Hess Patrick D
Actelion Pharmaceuticals, Allschwil, Switzerland.
J Am Assoc Lab Anim Sci. 2012 Mar;51(2):231-8.
Here we evaluated the ability of a new, dual blood-pressure telemetry transmitter to simultaneously measure pulmonary and systemic blood pressure and the electrocardiogram in rats. The transmitter was implanted in normotensive and monocrotaline-induced pulmonary hypertensive Wistar rats, with sensing catheters placed in the pulmonary artery (channel 1) and descending aorta (channel 2). Biopotential electrodes were positioned to record an apex-based lead II electrocardiogram. Pulmonary and systemic arterial blood pressure and electrocardiographic waveforms were recorded between 2 and 12 wk after implantation of the transmitter. During this period, pulmonary arterial pressure progressively increased in monocrotaline-treated compared with saline-treated rats. The pharmacologic response of rats to reference compounds was measured by using the transmitter to validate the technique and to evaluate the ability of the device to transmit changes in blood pressure and the electrocardiogram. Validation against 2 Millar high-fidelity blood-pressure catheters confirmed the accuracy of the blood pressure data recorded with the transmitter. In addition, local tolerance of the associated catheters was confirmed by histologic examination.
在此,我们评估了一种新型双血压遥测发射器同时测量大鼠肺动脉压、体循环血压和心电图的能力。将该发射器植入正常血压和用野百合碱诱导的肺动脉高压Wistar大鼠体内,传感导管分别置于肺动脉(通道1)和降主动脉(通道2)。放置生物电位电极以记录基于心尖的II导联心电图。在发射器植入后2至12周记录肺动脉压、体循环动脉血压和心电图波形。在此期间,与盐水处理的大鼠相比,野百合碱处理的大鼠肺动脉压逐渐升高。通过使用该发射器测量大鼠对参考化合物的药理反应,以验证该技术并评估该设备传输血压和心电图变化的能力。与两根Millar高保真血压导管进行对比验证,证实了用该发射器记录的血压数据的准确性。此外,通过组织学检查确认了相关导管的局部耐受性。