Int J Sports Med. 2012 Dec;33(12):994-9. doi: 10.1055/s-0032-1312580. Epub 2012 Jul 12.
The main purpose of this study was to assess the inter-monitor reliability of the tri-axial GT3X Actigraph accelerometer over a range of physical activities (PA). This device collects motion data on each of the vertical (Y), horizontal right-left (X), and horizontal front-back (Z) axes and also calculates the vector summed value √X(2)+Y(2)+Z(2) known as 'vector magnitude' (VM). 8 GT3X accelerometers were worn at the same time by the same participant. Accelerometers were placed back-to-front, all facing forward and in sets of 4 securely taped together, attached to a belt and allocating each block above either left or right hip at waist level. Inter-monitor reliability was assessed during 6 conditions: rest, walking (4 and 6 km·h(-1)), running (8 and 10 km·h(-1)) and repeated sit-to-stand (40 times·min(-1)). The intra-class correlation coefficients were high for X, Y and Z axes (i.e., all ≥ 0.925) and for VM (≥ 0.946). In conclusion, we found good inter-instrument reliability of the GT3X accelerometer across all planes, yet our results also suggest that the X and Z axes do not provide further benefits over the 'traditional' Y-axis to assess the movement in typical PA.
本研究的主要目的是评估三轴 GT3X Actigraph 加速度计在一系列身体活动(PA)中的跨仪器可靠性。该设备在每个垂直(Y)、水平左右(X)和水平前后(Z)轴上采集运动数据,并计算矢量和值√X(2)+Y(2)+Z(2),称为“矢量幅度”(VM)。同一名参与者同时佩戴 8 个 GT3X 加速度计。加速度计从前向后放置,全部朝前,以 4 个一组用胶带固定在一起,系在腰带上,每块加速度计分别分配到左髋或右髋上方。在 6 种情况下评估跨仪器可靠性:休息、步行(4 和 6 km·h(-1))、跑步(8 和 10 km·h(-1))和反复坐立(40 次·min(-1))。X、Y 和 Z 轴(即,全部≥0.925)和 VM(≥0.946)的组内相关系数均较高。总之,我们发现 GT3X 加速度计在所有平面上的仪器间可靠性良好,但我们的结果还表明,X 和 Z 轴在评估典型 PA 中的运动时,并不比“传统”Y 轴提供更多的益处。