To assess the sensitivity of detection and accuracy of volumetry by manual and semi-automated quantification of artificial pulmonary nodules in an anthropomorphic thoracic phantom on low-dose CT.
Fifteen artificial spherical nodules (diameter 3, 5, 8, 10 and 12 mm; CT densities -800, -630 and +100 HU) were randomly placed inside an anthropomorphic thoracic phantom. The phantom was examined on 16- and 64-row multidetector CT with a low-dose protocol. Two independent blinded observers screened for pulmonary nodules. Nodule diameter was measured manually, and volume calculated. For solid nodules (+100 HU), diameter and volume were also evaluated by semi-automated software. Differences in observed volumes between the manual and semi-automated method were evaluated by a t-test.
Sensitivity was 100 % for all nodules of >5 mm and larger, 60-80 % for solid and 0-20 % for non-solid 3-mm nodules. No false-positive nodules but high inter-observer reliability and inter-technique correlation were found. Volume was underestimated manually by 24.1 ± 14.0 % for nodules of any density, and 26.4 ± 15.5 % for solid nodules, compared with 7.6 ± 8.5 % (P < 0.01) semi-automatically.
In an anthropomorphic phantom study, the sensitivity of detection is 100 % for nodules of >5 mm in diameter. Semi-automated volumetry yielded more accurate nodule volumes than manual measurements.
评估在低剂量 CT 上对人工肺结节进行手动和半自动量化时检测的灵敏度和准确性。
在人工胸部体模内随机放置 15 个人工球形结节(直径 3、5、8、10 和 12mm;CT 密度-800、-630 和+100HU)。体模用 16 排和 64 排多层 CT 进行低剂量检查。两名独立的盲法观察者对肺结节进行筛查。手动测量结节直径,并计算体积。对于实性结节(+100HU),直径和体积也通过半自动软件进行评估。手动和半自动方法之间观察到的体积差异通过 t 检验进行评估。
所有 >5mm 的结节和更大的结节的敏感性均为 100%,实性结节的敏感性为 60-80%,非实性 3mm 结节的敏感性为 0-20%。未发现假阳性结节,但观察者间的可靠性和技术间的相关性较高。与半自动方法相比,手动方法低估了任何密度结节的体积 24.1±14.0%,低估了实性结节的体积 26.4±15.5%(P<0.01)。
在人工体模研究中,直径 >5mm 的结节的检测灵敏度为 100%。半自动体积测量比手动测量更准确地获得结节体积。