Department of Radiology, Thomas Jefferson University, Philadelphia, PA 19107, USA.
Ultrasound Med Biol. 2012 Oct;38(10):1784-98. doi: 10.1016/j.ultrasmedbio.2012.06.013. Epub 2012 Aug 21.
The efficacy of using subharmonic emissions from Sonazoid microbubbles (GE Healthcare, Oslo, Norway) to track portal vein pressures and pressure changes was investigated in 14 canines using either slow- or high-flow models of portal hypertension (PH). A modified Logiq 9 scanner (GE Healthcare, Milwaukee, WI, USA) operating in subharmonic mode (f(transmit): 2.5 MHz, f(receive): 1.25 MHz) was used to collect radiofrequency data at 10-40% incident acoustic power levels with 2-4 transmit cycles (in triplicate) before and after inducing PH. A pressure catheter (Millar Instruments, Inc., Houston, TX, USA) provided reference portal vein pressures. At optimum insonification, subharmonic signal amplitude changes correlated with portal vein pressure changes; r ranged from -0.82 to -0.94 and from -0.70 to -0.73 for PH models considered separately or together, respectively. The subharmonic signal amplitudes correlated with absolute portal vein pressures (r: -0.71 to -0.79). Statistically significant differences between subharmonic amplitudes, before and after inducing PH, were noted (p ≤ 0.01). Portal vein pressures estimated using subharmonic aided pressure estimation did not reveal significant differences (p > 0.05) with respect to the pressures obtained using the Millar pressure catheter. Subharmonic-aided pressure estimation may be useful clinically for portal vein pressure monitoring.
利用 SonoVue 微泡(通用电气医疗,挪威奥斯陆)的次谐波发射来跟踪门静脉压力和压力变化的效果,在 14 只犬中使用门静脉高压的慢流或高流模型进行了研究。使用修改后的 Logiq 9 扫描仪(通用电气医疗,密尔沃基,威斯康星州,美国)以次谐波模式(发射频率:2.5 MHz,接收频率:1.25 MHz)运行,在诱导 PH 前后以 10-40%的入射声功率水平收集射频数据,每个水平有 2-4 个发射周期(重复 3 次)。压力导管(米勒仪器公司,休斯顿,德克萨斯州,美国)提供参考门静脉压力。在最佳照射条件下,次谐波信号幅度变化与门静脉压力变化相关;分别考虑 PH 模型或一起考虑时,r 值范围为-0.82 至-0.94 和-0.70 至-0.73。次谐波信号幅度与绝对门静脉压力相关(r:-0.71 至-0.79)。在诱导 PH 前后,次谐波幅度之间的差异具有统计学意义(p ≤ 0.01)。使用次谐波辅助压力估计估算的门静脉压力与使用 Millar 压力导管获得的压力没有显著差异(p > 0.05)。次谐波辅助压力估计可能在临床上对门静脉压力监测有用。